3.2.4
Continuity measurement
Range Resolution
Observation
No-load voltage
600.0
Ω
0.1Ω
The buzzer is actuated
at values <≈ 30Ω
approx. 0.8V
3.2.5
Diode test
Range
Resolution
Measurement uncertainties
No-load voltage
2.000V
0.001V
1.0%+5
approx. 2.0V
3.2.6
Capacitance measurement
Range
Resolution Measurement uncertainties
Remarks
6.000nF
0.001nF
2.5%+50
-
60.00nF
0.01nF
2.5%+8
600.0nF
0.1nF
2.5%+5
6.000μF
0.001μF
60.00μF
0.01μF
600.0μF
0.1μF
6.000mF
0.001mF
3.0%+5
60.00mF
0.01mF
4%+5
3.2.7
Frequency measurement
Range
Resolution
Measurement uncertainties
Notes
6.000kHz
0.001kHz
0.1%+3
-
60.00kHz
0.01kHz
Sensitivity
Range
Minimum input signal (sine wave)
10Hz~5.999kHz
6.00kHz~59.99kHz
600mV
60mV
-
6.0V
0.3V
3V
60V
2V
6V
600V
20V
60V
- 11 -
3.3 Environmental conditions
Environmental conditions
in use
in storage
Temperature
0°C to +50°C
32°F~122°F
-10°C to +60°C
14°F~140°F
Relative humidity (RH)
≤80% without
condensation
≤70% without
condensation
Altitude
<2.000m
up to 10,000m
3.4 Constructive characteristics
Housing
Rigid polycarbonate shell with moulded elastomer covering
Screen
LCD display
Dimension: L 41 x H 18mm
Dimension
H 130 x L 81 x P 24mm
Weight
150g (without battery)
3.5 Power supply
Batteries:
2x1.5V (AAA)
Mean battery life
>250 hours
Auto power off delay
After approximately 10 minutes without action on
the switch and/or on the keys
3.6 Compliance with international standards
Electric safety
Protection class
Compliant with standards IEC61010-1 and IEC61010-2-033
300V CAT III.
Class 2, double insulation
Electromagnetic
compatibility
Compliant with standard EN61326-1
Level of protection of
the housing
Housing: IP40 (according to the standard IEC60529)
- 12 -
4. MAINTENANCE
The instrument has no parts that can be replaced by personnel who are not trained and
approved. Any non-approved repair or other work, or replacement of a part by an
"equivalent", may severely compromise safety.
4.1 Cleaning
Disconnect the unit completely and turn the rotary switch to OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly
with a dry cloth or forced air.
Dry perfectly before putting back into use.
4.2 Replacement of the batteries
The
symbol indicates that the batteries are spent. When this symbol appears on
the display unit, the batteries must be replaced. The measurements and specifications
are no longer guaranteed.
To replace the batteries, proceed as follows:
1.
Disconnect everything connected to the device;
2.
Set the switch to OFF;
3.
using a cross-headed screwdriver unscrew the screw of the battery compartment
cover on the back of the housing and open the cover;
4.
Always replace both batteries;
5.
Close the cover and screw it back down.
4.3 Repair
Return the instrument to your distributor for any work to be done, whether under the
warranty or not.
If you have to ship the instrument, it is best to use its original packaging and to state as
clearly as possible, in a note attached to the equipment, the reasons for the transfer.
5. WARRANTY
The equipment is warranted against defects of materials or workmanship, in accordance
with the general terms of sale.
During the warranty period (1 year), the instrument must be repaired only by the
manufacturer, who reserves the right to choose between repairing it or replacing it,
entirely or partially.
- 13 -
If the equipment is sent back to the manufacturer, carriage is paid by the customer.
The warranty does not apply in the following cases:
1. Inappropriate use of the equipment or use with incompatible equipment;
2. Modifications made to the equipment without the explicit permission of the
manufacturer’s technical staff;
3. Work done on the device by a person not approved by the manufacturer;
4. Adaptation to a particular application not anticipated in the definition of the
equipment or not indicated in the user’s manual;
5. Damage caused by shocks, falls, or floods.
6. DELIVERY CONDITION
The DMM13 multimeter is delivered in a blister pack with:
2 1.5V batteries
a user manual in several languages
- 14 -