background image

6

Range

Resolution

Accuracy

600.0mV

0.1mV

+(1.0%+2) 

6.000V

0.001V

+(0.7%+3) 

60.00V

0.01V

+(0.7%+3) 

600.0V

0.1V

±(1.2%+3)

60Hz-1MHz

0.01 Hz/0.001 MHz

±(0.1%+5)

Notes

:

• 

Overload protection: 600Vrms (AC/DC). Input impedance: about 10MΩ 

• 

Display sine wave true RMS. Frequency response: 40Hz -400Hz 

• 

Max input voltage: 600Vrms. when the voltage >610V, “OL” symbol appears and 

the buzzer goes off. 

• 

Frequency Sensitivity about 300mV 

RESISTANCE

Range

Resolution

Accuracy

600.0Ω

0.1Ω

+(1.0%+2)

6.000kΩ/6000Ω 

0.1kΩ/1Ω

+(0.8%+2)

60.00kΩ

0.01kΩ

±(0.8%+2)

600.0kΩ

0.1kΩ

+(0.8%+2)

60.00MΩ

0.01MΩ

+(2.0%+5)

DIODE TEST

Range

Resolution

Remarks

0.001V

MP730007

 

Open circuit voltage: 4V, test current: about 1.5mA. 

MP730008

Open circuit voltage: 2.1V, test current: about 1mA.  

Silicon PN junction voltage: 0.5 - 0.8V.

OPERATING PARAMETERS

• 

Ambient temperature 23

º

 ± 5

º

• 

Relative Humidity: <75%

DC VOLTAGE

Range

Resolution

Accuracy

600.0mV

0.1mV

±(0.7%+3) 

6.000V/6000mV

0.001V/1mV

±(0.5%+2)

60.00V

0.01V

±(0.7%+3)

600.0V

0.1V

±(0.7%+3)

Note: 

Input Impedance: 10MΩ for all ranges

AC VOLTAGE

Note: 

• 

Measurement result = reading of resistor - reading of shorted test leads 

• 

Overload protection: 600Vrms 

Содержание MP730007

Страница 1: ...1 Model MP730007 and MP730008 Digital Multimeter ...

Страница 2: ...0 seconds For current values over 5A the wait period between each measurement must be longer than 15 minutes When a measurement has been completed disconnect the testing probes from the circuit under test Replace the batteries as soon as the low battery indicator appears on the display Remove dead batteries from the meter or if it is not going to be used for a long time Never mix old and new batte...

Страница 3: ...put voltage 600V AC DC the buzzer will continuously beep indicating measure range is at it s limit If the input current 10A AC DC buzzer will continuously beep indicating measure range is at it s limit Disconnect the probes from the circuit under test when changing the range If the Display shows OL it indicates the measured voltage is out of range In this case turn the range selector to a higher r...

Страница 4: ...asurement When there is no input such as in open circuit condition the instrument will display OL When making an in circuit measurement disconnect all power supplies and discharge all the capacitors When measuring low resistance the test leads will produce 0 1Ω 0 2Ω measurement error To obtain accurate measurement short the test leads and use REL function If the resistance when shorted is more tha...

Страница 5: ...nge selector to battery measurement position Insert the red test lead to VΩmA terminal black to COM terminal Connect the red test probes to and black to poles The LCD displays the battery status Notes Do not input over 60V DC or 30V AC Measuring Temperature MP730007 only Switch the range selector to temperature position After OL is displayed insert the K type thermocouple to the device terminals V...

Страница 6: ... 0 2 6 000kΩ 6000Ω 0 1kΩ 1Ω 0 8 2 60 00kΩ 0 01kΩ 0 8 2 600 0kΩ 0 1kΩ 0 8 2 60 00MΩ 0 01MΩ 2 0 5 DIODE TEST Range Resolution Remarks 0 001V MP730007 Open circuit voltage 4V test current about 1 5mA MP730008 Open circuit voltage 2 1V test current about 1mA Silicon PN junction voltage 0 5 0 8V OPERATING PARAMETERS Ambient temperature 23º 5º Relative Humidity 75 DC VOLTAGE Range Resolution Accuracy 60...

Страница 7: ... is only applicable for temperature less than 250ºC 482ºF CAPACITANCE Range Resolution Accuracy 9 999nF 0 001nF REL mode 4 10 99 99nF 0 01nF 4 5 999 9nF 0 1nF 4 5 9 999µF 0 001µF 4 5 99 99µF 0 01µF 4 5 999 9µF 0 1µF 4 5 9 999mF 0 001mF 10 DC CURRENT Range Model Resolution Accuracy 600 0µA MP730007 MP730008 0 1µA 1 0 3 6000µA MP730007 1µA 1 0 3 60 00mA MP730007 MP730008 0 01mA 1 0 3 600 0mA MP73000...

Страница 8: ... ELECTRICAL ELECTRONIC EQUIPMENT These symbols indicate that separate collection of Waste Electrical and Electronic Equipment WEEE or waste batteries is required Do not dispose of these items with general household waste Separate for the treatment recovery and recycling of the materials used Waste batteries can be returned to any waste battery recycling point which are provided by most battery ret...

Отзывы: