Chapter 3 Basic Measurement 

Before measurement, be sure to read this chapter. 

 

To avoid electric shock accident and short circuit, please operate the instrument as following: 

 

Do not test the voltage over 70 VDC 

 

Do not test the terminal-to-ground voltage over 70 VDC. 

 

Do not test AC voltage. 

 

 

If a voltage or current is applied beyond the rated range, the instrument will be damaged. 

 

After measuring the high voltage battery, you should put the test leads together and make 
them shored first, and then test the low voltage batteries. In this way, the power of DC 
cutoff capacitance can be removed, otherwise the overvoltage may damage the 
low-voltage battery. 

 

 

Internal battery resistance varies considerably depending on charge or discharge status. To 
increase measurement accuracy, make measurements under similar conditions (for 
example, a fully charged battery). 

 

Lead-acid batteries (measuring objects) have high levels of terminal resistance. For this 
reason, resistance values may differ between the case and the tip of the terminal. Be sure 
to connect the test lead to the terminals at a fixed location. 

 

Use the optional Clip Type Lead with Temperature Sensor to measure the battery 
temperature. Or, use a non-  contact thermometer, such as a radiation thermometer, for 
safety. 

 

Measurement may not be possible for insulated terminals, due to insufficient flow of 
current for measurement. In such a case, clean the terminal (remove the insulation) before 
measurement.   

 
 
 
 
 
 
 

3.1 Pre-operation Inspection

 

Содержание MP700499

Страница 1: ...PORTABLE BATTERY TESTER MP700499 MP700502...

Страница 2: ...Power On Off 18 2 5 Clock Function 19 Chapter 3 Basic Measurement 21 3 1 Pre operation Inspection 21 3 2 Setting the Measurement Range 22 3 3 Zero Adjustment 22 3 4 Retaining the Displayed Values 25 3...

Страница 3: ...o hold Function 44 6 3 Auto memory Function 44 6 4 Auto Power Save Function 45 6 5 Battery Level Warning 46 Chapter 7 Specifications 47 7 1 General Specifications 47 Chapter 8 Maintenance and Service...

Страница 4: ...nspect it carefully to ensure that no damage occurred during shipping In particular check the accessories panel switches and connectors If damage is evident or if it fails to operate according to the...

Страница 5: ...Carrying Bag 9803 USB Cable zero adjustment block 9363 B Pin Type test lead Lithium battery Battery Charger Chinese standard Option 9363 A Test Clip British standard European standard Battery Charger...

Страница 6: ...fe operating condition Before using the instrument be certain to carefully read the following safety notes Indicates very important message in this manual When the symbol is printed on the instrument...

Страница 7: ...lue This is usually the maximum display value In the instrument this indicates the currently used range rdg Reading or displayed value The value currently being measured and indicated on the measuring...

Страница 8: ...lly to ensure that no damage occurred during storage or shipping If you find any damage contact your authorized Multicomp Pro distributor or reseller Before using the instrument check that the coating...

Страница 9: ...may occur when the test leads are connected to batteries which can ignite any accumulated inflammable gases such as hydrogen Handling leads and cables When using the instrument use only the test line...

Страница 10: ...ected by lead or connector resistance because it uses the AC 4 terminal method to measure internal resistance Simultaneous display of resistance voltage and temperature Without changing functions this...

Страница 11: ...irtually any location because the pin can be inserted diagonally in hard to reach places Remote Control Switch optional for storing measurement values The Remote Control Switch makes it possible to ho...

Страница 12: ...alue Execute zero ajustment function Holds or cancels the measurement values Switch system setup interface Deletes the settings Stores display value to memory Selects a configuration setting or change...

Страница 13: ...Set voltage range Zero Adjustment Performing zero adjustment is according to the zero ajusting method required by specified test leads Start of the test 1 Connect the test leads to the battery 2 Read...

Страница 14: ...14 1 5 Dimensions...

Страница 15: ...e strap 1 Switch off the instrument and remove the test leads 2 Pass the neck strap through the attachments marked in red circle and fasten it in place with the buckles 3 Adjust the length of the neck...

Страница 16: ...attach the cover Poor performance or damage from battery leakage could result Observe the cautions listed below Do not use both old and new batteries or different types of batteries together Be carefu...

Страница 17: ...17 LR6 battery replacement Lithium battery replacement...

Страница 18: ...e four terminals DRIVE and SENSE 2 4 Turning the Power On Off Press the POWER key to turn the power on or off Check the clock settings when using the instrument for the first time Turning the power on...

Страница 19: ...e date 1 Press the SET key on the test page twice to switch to the system setting interface 2 Select relevant options Press the direction key to move the cursor position press the ENTER key to select...

Страница 20: ...stem Settings interface 2 Select the relevant options Press the direction key to move the cursor position press the ENTER key to select the time and then press the direction key to adjust the value at...

Страница 21: ...oltage battery Internal battery resistance varies considerably depending on charge or discharge status To increase measurement accuracy make measurements under similar conditions for example a fully c...

Страница 22: ...efore measurement 3 2 Setting the Measurement Range Set resistance and voltage measurement ranges as described below Resistance range 3 m 30 m 300 m 3 Voltage range 6 V 60 V Resistance range Press the...

Страница 23: ...ior to measurement Be sure to use the included zero adjustment block 9363 B when performing zero adjustment Be sure to keep the test lead shorted during zero adjustment 3 3 1 Shorting Methods for Vari...

Страница 24: ...away from any metal Be sure to use the included zero adjustment board when performing zero adjustment When the distance between the terminals on the battery measurement target is more than the distanc...

Страница 25: ...new test leads Try to clean the zero adjustment board 4 The instrument automatically begins obtaining correction values When the zero adjustment operation is complete 0 ADJ is lit up and the instrumen...

Страница 26: ...n the resistance value is it indicates that the test line is open or disconnected or that the current is unstable for other reasons and cannot be tested will also be displayed when the test line is no...

Страница 27: ...nt and then press the left and right keys to turn on and off the temperature measurement function 3 Press ENTER to store the setting of temperature measurement function and press ESC to return to the...

Страница 28: ...ditions can be set A buzzer sounds when the test result is not qualified in the range 4 2 Turning On the Comparator Function 1 Press SET on the test page Switch to the comparator interface 2 Press the...

Страница 29: ...ance 4 4 Set the comparator group number The instrument has the function of preserving 200 record groups 1 Press SET on the test page Switch to the comparator interface Press the direction key to move...

Страница 30: ...ress the direction key to move the cursor to the resistance range position Press ENTER to select the resistance range to be set and then press left and right to set the resistance range Press ENTER ag...

Страница 31: ...t and then press the left and right keys to set the voltage range Press ENTER again to complete Press ESC to return to the measurement interface Set the comparator resistance threshold 1 Press SET on...

Страница 32: ...to move the cursor to the lower resistance position Press ENTER to select the lower limit of resistance to be set press left and right keys to move the cursor and press up and down keys to set the lo...

Страница 33: ...asurement interface 4 6 Setting the Comparator Buzzer The buzzer can be enabled in accordance with comparison results when the comparator function is used The buzzer can be set to sound in the followi...

Страница 34: ...erface 2 Press the direction key to move the cursor to the signal Press ENTER to select the mode of message ringing that needs to be set and press the left and right keys to switch between qualified r...

Страница 35: ...ction is turned on If there are no measurement values is displayed and a comparator judgment cannot be performed Even when the power is turned off the comparator settings are saved and the comparator...

Страница 36: ...to a computer The structure of the internal memory is as follows Memory Structure Unit name 12 units Memory number 200 cells A 1 2 3 198 199 200 b 1 2 3 198 199 200 C 1 2 3 198 199 200 d 1 2 3 198 19...

Страница 37: ...available memory number is displayed 5 3 Canceling the Memory Function 1 Press the SET key on the test page Switch to the system Settings interface 2 Press the direction key to move the cursor to auto...

Страница 38: ...read out is also displayed You cannot select the number of the memory location in which no data is stored If no data has been stored is displayed in the memory number display area and the instrument...

Страница 39: ...ata 4 Press ENTER again to confirm that the data stored in the selected memory address is deleted 5 5 2 Clearing a cell 200 groups of data 1 Press FILE on the test page Switch to the file reading inte...

Страница 40: ...ted cell is deleted 5 5 3 Clearing All Data 1 Press FILE on the test page Switch to the file reading interface 2 Press to select the memory address unit Press to select a place 3 Press CLEAR 3 times t...

Страница 41: ...d to the computer 1 Connect one end of the 9803 USB communication cable to the personal computer and the other end to the PC interface above the MP700499 2 When the cable is connected USB small icon a...

Страница 42: ...tti Pass Fail Temp Test Date Time 1 1 5192 1 0000 2 0000 4 05856 3 60000 4 30000 PASS 21 0 2020 4 30 10 59 2 1 5170 1 0000 2 0000 4 05850 3 60000 4 30000 PASS 21 1 2020 4 30 10 59 3 1 5167 1 0000 2 00...

Страница 43: ...able the average function can be used The average number can be selected as 2 3 or 4 1 Press SET on the test page Switch to the system Settings interface 2 Press the direction key to move the cursor t...

Страница 44: ...ress the ENTER key to select whether to turn ON the auto save function and press the left and right keys to switch OFF ON Press ENTER again to complete Press ESC to return to the measurement interface...

Страница 45: ...y when the auto memory function is used together with auto hold function 6 4 Auto Power Save Function The auto power save function can be used to reduce power consumption of the instrument The instrum...

Страница 46: ...ress ESC to return to the measurement interface 6 5 Battery Level Warning The battery level of the instrument is displayed at the upper right corner of the screen Battery level indicator Battery statu...

Страница 47: ...ranges in all 3m 30m 300m 3 30 Auto and manual test mode in 5 level range Sampling Rate 1 meas sec the average number optional 20 meas sec 3 meas sec Temperature Range 10 C 60 C Accuracy 1 C Correctio...

Страница 48: ...guarantee 1 year Operating temperature and humidity 0 C to 40 C 80 RH no condensation Storage temperature and humidity 10 C to 60 C 80 RH no condensation Operating environment Indoor maximum altitude...

Страница 49: ...s are inserted incorrectly C Reinsert the batteries correctly Incorrect measurement values are displayed is displayed on the screen OF is displayed on the screen The test lead is not connected correct...

Страница 50: ...gnal The instrument detects the internal inductive voltage if the voltage rises to a certain height will produce the wrong test signal Reducing the length of the test line especially the length of the...

Страница 51: ...the test line is very small which can be ignored In the instrument using a synchronous detection system the internal resistance is divided into impedance and reactance showing only the actual part if...

Отзывы: