First Use of the MEA Amplifier
3.1 Welcome to the MEA Amplifier
Raw data from up to
60 electrodes
of a
microelectrode array
(
MEA
) is amplified
by
60 channels
of
filter amplifiers
that are built very small and compact using SMD
(Surface Mounted Devices) technology. The small-sized MEA amplifier combines the
interface
to the
MEA probe
with the
signal filtering
and the
amplification
of the signal.
The compact design reduces line pick up and keeps the noise level down.
The
MEA sensor
is placed directly into the small-sized MEA
amplifier
. When the amplifier
is closed, the
contact pins
in the lid of the amplifier are pressed onto the MEA
contact pads
.
The very close location of the amplifier to the MEA sensor is very favorable concerning a high
signal-to-noise ratio.
MEA amplifiers can be ordered with different
gain
and
bandwidth
configurations by the user's
choice. For example, typical pass bands would be 1 to 300 Hz for recording field potentials and
300 to 3000 Hz for recording action potentials. It is also possible to use a broadband amplifier
and filter the data with the digital filter of the free MC_Rack program. Gain settings from
100 to 5000 are possible.
The amplifier is connected to the
data acquisition
computer via a single
68-pin MCS standard
cable
. The analog output signals of the MEA amplifier are then acquired and digitized by the
MC_Card (MEA-System) or your custom data acquisition system.
MEA amplifiers have an
integrated heating system
for controlling the
MEA's temperature
.
The desired temperature can be easily programmed with a temperature controller. You will not
need an incubator during recording all environmental conditions are reliably controlled directly
in the MEA dish.
You can then simply
ground
bad or unwanted electrodes by toggling small switches on the
amplifier. You can also connect a stimulus generator to any electrode(s) for
stimulation
.
Amplifier inputs are grounded during stimulation to prevent a saturation of the amplifier.