Grounding the bath
You can ground the bath with a plain silver wire or with an Ag/AgCl pellet. Only the Ag/AgCl
electrode provides a stable intrinsic potential. In practice, the plain silver wire will be good
for 90 % of applications. If you do not achieve satisfying results with the plain silver wire,
try an Ag/AgCl electrode.
If you use MEAs with internal reference electrode, you can use the reference electrode for
grounding the bath. This has the advantage that you can keep the culture chamber closed
and sterile (for example, with MEA-MEM semipermeable membranes).
Important: If you use a MEA
with internal reference electrode
, you have to
connect pin 15
to ground
with the provided
cable
. Setting electrode 15 to ground in the MEA_Select program
is
not
sufficient for grounding the bath.
1. Attach the provided
silver wire
or
pellet
to the amplifier's
ground
and place it into the
bath
,
near the rim of the culture chamber. If you use a MEA with
internal reference electrode
,
connect the ground to the
reference electrode socket (pin 15)
with the provided connector.
Make sure that the orientation of the MEA inside the amplifier is appropriate.
2. Replace the shielding.