mtts Impala - V1 Скачать руководство пользователя страница 12

V-0-02-EN

Impala User manual 

22

23

Alarm system

Alarm system

Alarm

Cause/Ventilator Response

Corrective Action

High rate Adjustment of Max the total of 

breath per minute too low

Re-adjust Max the total of breath per minute

Adjustment of the inspiratory sensi-

tivity level too low

Adjust trigger sensitivity according to the pa-

tient

Patient hyperventilating

Silence the alarm and call for a medical team if 

the symptoms persist
Check for auto-cycling and adjust inspiratory 

sensitivity, manage leaks or drain condensation 

from patient circuit

Defective flow sensor

Have a qualified technician replace the defective 

component(s) and call your customer service 

representative

Low rate

Adjustment of Max the total of 

breath per minute too high

Re-adjust Max the total of breath per minute

Adjustment of the inspiratory sensi-

tivity level too high

Adjust trigger sensitivity according to the pa-

tient

Defective flow sensor

Have a qualified technician replace the defective 

component(s) and call your customer service 

representative

High Vti

Adjustment of the Max Vti level too 

low (for PSV, CPAP, A/C PC, SIMV 

PC, SIMV VC modes).

Note: Always consult the clinical before chang-

ing PEEP, FiO2, pressure, volume or rate set-

tings.

Modify the Max Vti level

Flow sensor not calibrated properly

Calibrate the flow sensor

Low Vti

Adjustment of Min Vti too high or f 

pressure level not enough to reach 

the volume required (for PSV, CPAP, 

A/C PC, SIMV VC, SIMV PC modes)

Modify the Min Vti level or the pressure level ac-

cording to the physician’s prescription

Patient circuit obstructed or discon-

nected

Clean, unblock, and/or re-connect the patient 

circuit

Flow sensor not calibrated properly

Calibrate the flow sensor

Low FiO2 Not enough oxygen delivered by the 

ventilator

Check oxygen source. 

Replace oxygen tank if it’s empty.

High FiO2 Oxygen sensor not calibrated prop-

erly

Calibrate the oxygen sensor

The Impala Ventilator will generate visual and audible alarm signals to notify you of

alarm conditions detected.

An alarm condition is indicated by:

•  Audible alarm tone

• 

Visual alarm indicator (flashing parameter button and power button)

In order to verify the operation of auditory and visual alarms, perform the following 

procedure:

1. On the main screen, disconnect flow sensor and start ventilation

2. Check pressure audible alarm sounds

Alarm system test

Alarm silence

Audible alarms may be suspended, while visual alarms may not. The alarm silence 

button is used to temporarily silence audible alarms for 2 minutes. When this but-

ton is pressed, its color turns red indicating audible alarms are silenced. The alarm 

sounds are can be restored by press the button again.

List of alarms

Alarm

Cause/Ventilator Response

Corrective Action

Low battery

The AC (mains) power source 

cut-off 

and internal battery capacity is 

less than 30%.

Reconnect the device to an AC power outlet

High pressure

Adjustment of Max PIP too 

low (only for A/C VC, SIMV VC 

modes)

Note: Always consult the clinical before 

changing PEEP, FiO2, pressure, volume or 

rate settings. Increase the Max PIP threshold

Airway obstruction

Check patient’s trachea and clear the ob-

struction. If the filter is obstructed, replace 

the filter

Coughing or other high-flow 

exhalation efforts

Treat patient’s cough. Silence the alarm, if 

necessary

Patient inspiratory resistance or 

compliance changes

Have physician determine if ventilator set-

tings are appropriate for the patient.

Содержание Impala - V1

Страница 1: ...Impala User manual User Manual Effective date July 2020 Issue Number 00 Version 02 EN Issue Date June 2020 www mtts asia com Impala ventilator Model Impala V1...

Страница 2: ...manual Contents Company Information Introduction Device Description Warnings Setup Operation Alarm System Cleaning Maintenance Specifications Explanation of Symbols Warranty Policy 04 05 06 09 11 16...

Страница 3: ...tnam Tel 84 24 3766 6521 Fax 84 24 3718 8050 Email assistance mtts asia com www mtts asia com PROMETEO SRL Via Mantova 9 38122 Trento Italy This user manual is intended for health care professionals T...

Страница 4: ...current ventilation mode Touch this button to change mode 6 Standby Button Enable Disable Standby mode 7 Alarm Silence Button Touch this button to mute auditory alarm for 2 minutes 8 Setting Button T...

Страница 5: ...k of the device The presence of electro surgical devices shortwave or microwave equipment near this device could lead to electrical interference that would negatively affect the function of the device...

Страница 6: ...ay compromise its functional ity and system performance leading to a possible operator or patient hazard Performance is not guaranteed if an item labeled as single use is reused An alternative means o...

Страница 7: ...ratory tube 4 Attach exhalation valve control tube to the control port in front of the device 5 Insert a flow sensor into the breathing circuit in front of the patient connec tion 6 Attach the blue an...

Страница 8: ...ching the compressed oxygen will not waste any oxygen The machine will not use any compressed oxygen until you increase the FiO2 setpoint The device can work with both single limb and double limb incl...

Страница 9: ...the sensor during cali bration process The operator will be asked to flip the sensor during calibration process 16 17 Operation Set patient information After turning on a window automatically appears...

Страница 10: ...ifferent modes are compatible This reduces the risk of discom fort and harm to the patient Ensure another source of ventilatory support is available during the calibra tion and testing The patient mus...

Страница 11: ...limits Check system tightness Adjust less frequenly used parameters Adjust alarm sound loudness The Setting screen and available features may vary depending on the soft ware version Turn the device o...

Страница 12: ...or re connect the patient circuit Flow sensor not calibrated properly Calibrate the flow sensor Low FiO2 Not enough oxygen delivered by the ventilator Check oxygen source Replace oxygen tank if it s...

Страница 13: ...maximum 2 in wa ter Dry all the surfaces after cleaning with a clean soft cloth or paper towel Ensure that no part of Impala Ventilator is immersed in any cleaning liquid or cleaning solution Do not...

Страница 14: ...rts that must be replaced or serviced during the lifetime of the device These parts include Component Approx Duration of Use Maintenance Information Air filter 6 months Clean or replace Oxygen sensor...

Страница 15: ...ressure PIP Fraction of Inspired Oxygen FiO2 Frequency Respiratory rate Support pressure Tidal volume Vt 0 25 cmH2O 5 60 cmH2O 21 100 2 60 bpm 5 60 cmH2O 25 2000 ml Monitoring parameters Specification...

Страница 16: ...manufacturer Date of manufacture Keep out of direct sunlight This device should be disposed of separately from normal household waste so that components can be recycled Refer to User Manual before op...

Страница 17: ...new materials and used materials equivalent to new in performance and reliability MTTS may repair or replace MTTS products a with new or previously used products or parts equivalent to new in performa...

Страница 18: ...stall software patches run MTTS diagnostics and utilities Implement temporary procedures or workarounds provided by MTTS while MTTS works on permanent solution Cooperate with MTTS in attempting to re...

Страница 19: ...ER AND SERVICES Co LTD House No 26 Alley 41 An Duong Vuong Street Tay Ho District Hanoi City Vietnam Tel 84 24 3766 6521 Fax 84 24 3718 8050 Email assistance mtts asia com www mtts asia com PROMETEO S...

Отзывы: