d) Channel Goal Settings: Set the microvolt goal for each channel; this can also be
set or adjusted in the Train module by assessing MVC in real time. Goal can also
be modified by typing into the text bar (tap on the number before uV to type).
Note
: 1000-1500uV is a good starting point to assess patient capability if setting
goal manually
Learn more about goal setting here.
e) Time Settings: Modify total, contract, and relax time as desired. Both contract
and relax times must be set or you will be notified of an error. Number of reps is
calculated as [total time]/[flex time + relax time]. Every training session will begin
with the relax period of the rep.
f) Ch1 Audio Select: Toggle Ch 1 Audio ON to hear cues for relax and contract
periods as well as a “success tone” when the contraction reaches the green
zone. Audio is only reflective of Ch 1 activity. Toggle Audio OFF to mute. If using
audio, ensure mobile device has media volume turned on.
11)
Save settings by tapping SAVE in upper right corner. Return to the Home screen by
tapping
“BACK” in upper left or
m
in lower left. If settings have not already been
saved, accept
“save settings” prompt upon exiting.
12)
From the Home screen, tap to enter Train.