Operating Input Calibration
Operating calibration in order of items and
calibration points in Table 10-3
:
Table 10-3
Item
No.
Input Range
Calibration Point
1
DCA/20mA
0 : 00.000mA
FS : 19.000mA
2
DCV/28V
0 : 00.000V
FS : 19.000V
Current Measurement Calibration
1) In power-off state, press the keys
〔
OUTPUT
/ INPUT
〕
and
〔
power
〕
simultaneously to
power on, and release the
〔
OUTPUT /
INPUT
〕
key only after the LCD displays all
the contents. The Instrument enters into the
calibration mode, and the symbols ‘CAL 0’
appear in the LCD.
2) Press the
〔
OUTPUT / INPUT
〕
key to
select measurement function , and the LCD
displays ‘INPUT ’,‘mA ’;
3) Connect to the Instrument as shown in
Figure 10-2;
T R IG
O U T
A U X
S C O P E
L o
N O R M A L
H i
S ta n d a rd S o u rc e
( 5 5 2 0 A )
Figure 10-2
4) Set the standard source to an appropriate
range;
5) Set the output of standard source as the
value given in Table 10-3 and when the
output
stabilized
,
press
the
34
35