2) Press the key
〔
OUTPUT / INPUT
〕
to
select output function , and the LCD displays
‘OUTPUT ’,‘mA ’;
3) Connect to the Instrument as shown in
Figure 10-1;
4) Set the digital meter to an appropriate range;
5) When the output stabilized , press the
key
〔
〕
/
〔
〕
and
the
key
〔
〕
/
〔
〕
to adjust the calibrator to a
value in identity with the reading of the digit
meter;
6) Press the key
〔
100%/START
〕
, and the
LCD displays ‘SAVE ’, denoting that the
calibrated point has been stored;
L O
B Z -1 0
H I
K E IT H L E Y 2 0 0 0
Figure 10-1
7) Press the key
〔
STEP/AUTO
〕
and the LCD
displays
‘
CAL 0 FS
’
. With the output
stabilized, repeat the operation of steps 5
and 6;
8) Press the key
〔
STEP/AUTO
〕
and the LCD
displays
‘
CAL 0 FS
’
. With the output
stabilized, repeat the operation of steps 5
and 6.