User Manual - PolyXeta
®
2 Detector with HL Sensor
for Refrigerant Gases - May_2020
Page
24
PolyXeta
is a registered trademark of MSR
Made in Germany
GAPX2_Freon_E_2020_05_18
Phone: +49 8531 9004-0 Fax: +49 8531 9004-54
Specification subject to change without notice
MSR-Electronic GmbH, Buergermeister-Schoenbauer-Str. 13, D 94060 Pocking
www.msr-electronic.de
Table 10.2: Sensor specifications for refrigerant & Freon gases
Sensor
head
Target gas
Test gas
Admissible range for calibration
Reaction
time
Respons
e time
↑
t
90
Respons
e time
↓
t
90
Run-in
time
Flow
rate
Warm-up
time
Meas.
range
Gas type
Range
Value indication
4
–
20 mA signal
SX1-
Freon
group
Gas
Zero
Gain
Zero
Gain
Test gas
(mA)
Sec.
Sec
Sec
h
ml/min Sec.
S2061
20
–
2000 ppm
FR02
R23
1000 ppm
0
–
100
ppm
800
–
1200
ppm
3.8
–
4.2
± 0.2 mA
of set-
point
value
180
> 120
> 120
168
500
300
S2063
FR03
R1234yf
S2064
FR04
R123
S2070
FR06
R22
S2077
FR07
R134a
S2080
FR08
R407c
S2020-01
0
–
50 % LEL
R32
25 % LEL
0
–
2 %
LEL
23
–
27 %
LEL
S2020-02
R455a
25 % LEL
S2020-03
R455b
25 % LEL
S2020-04
R1234yf
25 % LEL
S2020-05
R1234ze
25 % LEL
S2125-C 0
–
1000 ppm
NH
3
168
500
Max. 90
S2125-F 0
–
10,000ppm
NH
3
168
500