5.10 Service Counters
All MSI’s RF Linked Scales and Dyna-Links maintain two service counters for safety. The
first counter (LFCnt) counts lifts above 25% of capacity. The second counter (OLCnt)
counts the number of times the RF Linked Scale has been overloaded. These counters
serve to warn the user to inspect the load train after a number of overloads or a long term
frequency of high capacity lifts. Service counters can only be reset by the factory. The
power up routine will be interrupted when the lift counter exceeds 16383 lifts or the
overload counter exceeds 1023 overloads. Push any key to continue operation.
Feature only available on
MSI-8000
Software release 2.00 and above. Service counters are
available on the Scale/Dyna-Link front panel Test function.