8
2-6 Stage
①
Turn the coarse focusing knob to lower the
stage.
②
Open the spring finger
②
on the slide holder,
slide one or two sample slides from the front into
the stage.
③
Push the slide in as far as you can, and then gently release the trigger
The slide size should be 26 × 76mm, 0.9 to 1.2mm in thickness, and 0.17mm in thickness.
When observing a large sample, remove the clip and put the sample directly on the stage.
2-7 Adjust the tension of the x-axis and y-axis knobs
Hold the x-axis knob ①, screw the y-axis knob ②
upward to expose the knob rod.
1
、
Turn the x-axis adjusting knob
③
or y-axis adjusting
knob
④
clockwise (arrow direction) to increase the
tension, and turn the y-axis adjusting knob
④
counterclockwise to decrease the tension.
If the tension is adjusted too tightly, the
clucking sound can be heard when the stage rises
or falls, or the accuracy of the stage stop is reduced.
Rotating stage
1、Slightly loosen the table fixing knob ①.
2、Using the fixed knob of the loading platform,
the platform can rotate clockwise or
anticlockwise.
It is normal to hear and feel a click
during rotation