background image

- 5 -

 - 4 -

Water vapour detector tubes

20 to 160 mg/m3 water vapor at a flow rate of 1 l/min 
and  a  measuring  time  of  5  min.  (required  sampling 
volume  5  l)  or  5  to  70  mg/m3  water  vapor  at  a  flow 
rate  of  1  l/min  and  a  measuring  time  of 
10 min. (required sampling volume 10 l).

See also specific instructions for use of the 

KWIK DRAW

/ MSA AUER detector tubes H O-HP.

2

All values above are based on air not under pressure
and  at  room  temperature  and  normal  pressure.

5. Operating Instruction

5.1 Preaparing for Measuring 

5.2 Measuring on Compressed Air Cylinders

When  the  test  device  is  clean  and  dry  flushing  for 
2  min.  is  sufficient.  If  the  condition  of  device  is  not 
known,  or  if  during  a  previous  measurement  higher 
values were found than are permissible according to 
EN  12021,  flushing  at  2  l/min.  must  be  continued. 
T h i s   m u s t   b e   d o n e   u n t i l   2   w a t e r   v a p o r  
measurements,  made  at  an  interval  of  approx.  15 
min.,  give  identical  results  within  the  permissible 
measuring accuracy range. It then can be assumed 
that  the  test  device  is  sufficiently  clean  for  making 
measurements.                               

    .

Liquid/ water (droplets) can only be removed from 
the  test  device  through  long  flushing  with  dry,  if 
necessary  warm  air.                                            . 
                                                              .
Observe  arrow  on  the  detector  tube.  The  arrow 
points in the flow direction of the test air; therefor,
after inserting the tube, the arrow must point away
from  the  test  device.                                .

Clean  valve  of  compressed  gas  cylinders  or   
charging  connections  of  compressor  with  brief 
burst  of  air.                                        .
Chose  and  connect  the  adapter  suitable  for  the 
prescript  maximum  pressure  of  the  compressor
charging panel (300 or. 200 bars).                       .
Connect  pressure  reducer  to  compressed  gas 
cylinder, or to charging connection of compressor 
using  the  adapter.                            .
Insert and firmly fit  the nozzle of manifold on the 
pressure  reducer  as  shown  in  fig.  2           

    .

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

.

-

-

-

-

-
-

-

-

-

Break  off  the  detector  tube  tips  at  air  exit  side 
(see  fig.  5)                                             

      .

Put  detector  tube  into  the  tube  holder,  inserting 
broken-off  end  tightly  into  holder  side  away  from 
test  device.                                       

    .

Break  off  other  end  of  the  detector  tube  (air  entry 
side)  and  immediately  insert  open  tube  tightly  into 
other  end  of  detector  tube  holder  (close  to  test 
device).  (see  fig.  6).                                      .                   
Immediately  thereafter  start  the    timer  to  the  time 
designated  for  the  type  of  detector  tube. 
Remove  the    detector  tube  of  CO,  CO2  and    H2O 
after  5  minutes  from  the  tube  holder  and  take  the 
readings  from    the  printed  scale  on  the  tube. 
Remove the   oil detector tube from the tube holder of  
test  device  after  24  minutes  and  take  the  readings 
from the printed scale on the tube. 

5.3 Measuring on compressors

Switch on the compressor
Wait until 200 or 300 bar are reached
Switch off the compressor
Slowly open valve on compressor
Flush test device as scripted in para. 5.2

The  water  content  that  is  measured  in  air  from  a 
compressor  only  corresponds  to  the  water  content  of 
compressor gas cylinders simultaneously charged, or 
cylinders  charged  immediately  after  measuring,  if  a 
pressure retaining valve is built into the compressor at 
the  exit  side  of  the  drying  cartridges.  A  prerequisite, 
however, is that the cylinders to be charged are dry.

The pressure retaining valve assures that the pressure 
inside  the  compressor  drops  only  insignificantly  from 
the  maximum  pressure  of  300  or  200  bar,  when 
cylinders are being charged. With compressor that has 
no  pressure  retaining  valve  behind  the  filters,  water 
from  the  drying  cartridge  can  be  released  to  the 
sudden decompression when the charging connection 
is  opened.  The  air  has  then  higher  water  content  as 
was measured on the compressor charging panel. 

   .

MSA AUER compressors normally have a pressure
retaining valve. 

5.4 Permissable Test Results:

CO Contents
CO  Contents

2

Oil Contents 
Water Contents

max. 15 ppm
max. 500 ppm
max. 0.5 mg/m³
max. 50mg/m³ rel. to 200 bar)
max. 35 mg/m³ (rel. to 300 bar
max. 25 mg/m³ (if sampled from
compressor) 

-

-

-

-

-

-

-
-
-
-
-

Fig. 2: Inserting manifold  nozzle to 
          Pressure reducer coupling 

Fig. 5: Breaking off detector tube tips

Fig. 6: Inserting detector tube into detector tube
           holder

Slowly  open  cylinder  valve.                                                . 
The flow rate of air through each port of manifold 
can  be  confirmed  if  necessary  by  using  the  flow 
meter  provided with the kit. (To measure the flow 
rate connect  tubing of flow meter to the desired 
manifold  port-  Fig:3)                                                  .
Insert  the  detector  tube  holders  in  the  manifold 
ports-  Fig:4                                                    .
Flush  device  for  sufficient  time  to  make  the  test 
device free from other gases if any.                     .

                 

 .

Fig. 3: Connecting flow meter to 
           manifold port 

Fig. 4: Inserting detector tube holder
           in manifold port

Отзывы: