38
Cleaning and Disinfecting
G1 Powered Air-Purifying Respirator (PAPR)
US
(1) Use the instructions on the label to prepare a solution of Confidence Plus Germicidal Cleaner
(PN 10009971) in a bucket or sink.
NOTE:
The head harness can be cleaned as part of the facepiece or removed to be cleaned sepa-
rately.
(2) Submerge the facepiece in the solution of Confidence
Plus Germicidal Cleaner for a minimum of 30 seconds.
If necessary, use a moist sponge or soft brush to clean the
facepiece.
(3) Rinse the facepiece and components thoroughly in clean
water that is not more than 110°F (43°C) and is preferably
running and draining.
(4) To clean and rinse the pressure-demand exhalation valve,
use a blunt object to push in on the stem and flush the
valve with clean water.
(5) Let the facepiece air-dry. Do not put parts near a heater or in direct sunlight to dry.
(6) Operate the exhalation valve manually to make sure that it operates correctly.
(7) Make sure there is no water, moisture, or dampness on or in the facepiece before returning it to
service.
(8) Do a negative pressure seal test before returning the facepiece to service.