Kit Contents
•
Charging cradle
•
Power supply
2
Use
ALTAIR 4 Charging Station
For use with ALTAIR 4, ALTAIR 4XR Gas Detectors, and the ALTAIR Pump Probe
ALTAIR 5 Charging Station
For use with ALTAIR 5, ALTAIR 5X
2.1
Charging Times for Depleted Packs
•
ALTAIR 4/ALTAIR 4XR: 4 hours in normal, room temperature environments.
•
ALTAIR 5/ALTAIR 5X (PID/IR): 6 hours in normal, room temperature environments.
Minimum and maximum ambient temperature to charge the device is 10 °C (50 °F) and 35 °C (95 °F), respectively. For best
results, charge the device at room temperature 23 °C (73 °F).
2.2
Charging ALTAIR Gas Detectors
WARNING!
Risk of explosion: Do not recharge device in hazardous area.
Failure to follow this warning can result in serious personal injury or death.
WARNING!
•
Do not use the charger in a hot or wet environment.
•
Devices with a temperature below 10 °C (50 °F) or above 35 °C (95 °F) must stabilize for one hour at room temperature
before attempting to charge.
Failure to follow these warnings can result in serious personal injury or death.
WARNING!
The charger must be disconnected for the device to operate.
Failure to follow this warning can result in serious personal injury or death.
NOTE:
The charger may become hot while charging. This is not a malfunction.
NOTE:
Disconnect the charger from an external power source if not in use.
Insert the gas detector tightly into the charger as shown.
•
The battery symbol
will scroll through a progressively increasing number of
segments.
•
The charge LED will be red until 90% of full charge is obtained.
•
The battery symbol will remain fully illuminated, and the charge LED will be green
while the battery is trickle charged to its full capacity.