OPERATING MANUAL OXYGEN MONITORING SYSTEM
OMS 420
42
12
Appendix
12.1
Error codes
02 Configuration Error, new adjustment required
04 Flash Error, new adjustment required Please contact your dealer (manufacturer).
08 Unit to warm, ambient temperature too high!
01 Heating current<0.5A, O2 sensor not o.k.
32 Heating current<0.1A, CO sensor not o.k.
64 Heating current<0.1A, CO sensor not o.k.
128 Calibration error, calibration factor cannot be set.
Test gas cylinder is empty, filter dirty or clogged.
Calibration offset value cannot be set.
Instrument air supply is not available.
At the same time if several errors line up, then the error codes are to be added.
Her one example:
33 Heating current<0.5A, O2 sensor not o.k. and
Heating current<0.1A, CO sensor not o.k.