OPERATING MANUAL OXYGEN MONITORING SYSTEM   

OMS 420

 

 

33

 

9.7.5

 

O2 cell setup (by factory only) 

If there is no calibration gas available and the cell must be replaced, this function permits changing a cell without calibration. This 
can only be done if the cell factors have been measured at MRU and delivered together with the spare sensor. 

 

  

  

  

  

  

  

 

PREV or NEXT 

move cursor 

EDIT 

 

change value by means of UP/DOWN and POS 

OK  

save 

values 

BACK 

 

return to calibration menu

 

1) 

O2 cell setup requires input (edit function) of O2 factor and O2 offset values delivered with the spare sensor 

2) 

Temperature compensation of  20.9 % O

2

 is required when oxygen value at air is changed too much because of gas or 

flange temperature change. The difference of air oxygen measurement at lower temperature to measurement at higher tempera-
ture will be edit as “Temp.offs” value. 

NOTE: 

This procedure should be followed by new calibration using calibration gas cylinder.  

Only authorized persons or manufacturer’s trained technicians are authorized to perform adjustments on the OMS420. 

 

9.7.6

 

CO cell setup (by factory only) 

If there is no calibration gas available and the cell must be replaced, this function permits changing a cell without calibration. This 
can only be done if the cell factors have been measured at MRU and delivered together with the spare sensor. 

 

 

 

   

PREV or NEXT 

 

move cursor 

EDIT 

 

 

change value by means of UP/DOWN and POS 

OK 

  save 

values 

BACK 

 

 

return to calibration menu

 

CO cell setup factor and CO offset setup factor will be edit using the values delivered together with the spare sensor.  

 

Set point of heat resistor [Ohm] must be also carried out according to § 6.7.12 
NOTE: 

This procedure should be followed by new calibration using calibration gas cylinder.  

Only authorized persons or manufacturers trained staff are allowed to perform adjustments on the OMS420!

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание OMS 420

Страница 1: ...OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...packing material for use if the analyzer must be shipped in the future The products described in this manual are subject to continuous development and improvement and it is therefore ac knowledged tha...

Страница 3: ...OMS420RT 63467RT 14 7 3 3 High temperature model OMS420HT 63467HT 15 7 4 Principle of flow guidance tube 16 7 5 Transmitter electronics 17 7 6 Manual calibration 18 7 7 Recommended calibration gas 18...

Страница 4: ...CO cell setup by factory only 33 9 7 7 Adjustment O2 cell amplifier by factory only 34 9 7 8 Adjustment 4 mA 20 mA factory only 35 9 7 9 Calibration AIR 36 9 7 10 Calibration AIR GAS 36 9 7 11 CO cor...

Страница 5: ...should be stored in a safe but accessible place This instruction manual should be delivered to the end user immediately upon delivery NOTICE It is prohibited to transfer part or all of this manual in...

Страница 6: ...ually over the LEL low explosion level Power must always be provided to the probe even during boiler shut down to prevent the formation of condensate which can damage the sensors Do not use the probe...

Страница 7: ...y micro controller based transmitter electronics and converted into a standard 4 20 mA signal lin earized for oxygen in the range of 0 25 The expected lifetime of this sensor is about 5 years under no...

Страница 8: ...art below as these gases are oxidized at the hot sur face of sensor The cell voltage is measured by micro controller based transmitter electronics and converted into a standard linearized 4 20 mA sign...

Страница 9: ...combustion Optimum combustion conditions are achieved by decreasing the amount of excess air in the stack gas to the point where combus tibles start to increase See combustion diagram below The absolu...

Страница 10: ...f combustion optimization which in turn represents savings in fuel 7 2 OMS420 features The main features of OMS420 system are Compact reliable and rugged industrial design Special reference air not re...

Страница 11: ...to extract the sample from the tip of ceramic tube It can be used for clean flue gas temperatures up to 3 100 F 1 700 C 7 3 1 Compact model OMS420 61417 This model shall be used at site with low heat...

Страница 12: ...N65 SS316Ti material for temperature below 1 200 F 650 C AISI300 material for temperature below 1 800 F 1 000 C mounting flange assembly supplied by user detector head with junction box for sensors O2...

Страница 13: ...cable between junction box and transmitter pneumatic device option for automatic calibration back purge is not available for this model The main differences between the remote transmitter model and t...

Страница 14: ...behind the sensors ambient air for reference gas diffuses through another filter screen for dust protection and flushes the backside of sensors the reference side of the detector Illustration for comp...

Страница 15: ...65 NEMA4X that contains Printed circuit board with u processor PCB MAIN Backlit graphic display and dust proof keypad Printed circuit board for the sensor connection PCB OMS Electrical connector for p...

Страница 16: ...to calibration inlet port of unit The procedure of calibration is described in chapter 6 7 7 7 Recommended calibration gas For automatic calibration or manual calibration see procedure in chapter 6 7...

Страница 17: ...OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420 19 8 Software flow chart www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 18: ...ge has been properly insulated to prevent gas condensation Temperature Check mounting flange temperature min 160 F 70 C and max 300 F 150 C Check ambient temperature of transmitter for max 100 F 60 C...

Страница 19: ...O2 value wet The O2 display is calculated to wet dry The O2 display is calculated to dry The gas humidity can be set by SET to Set gas humidity 4 20 mA 02 setting range of analog O2 output 4 20 mA CO...

Страница 20: ...ng on flange temperature 2 V O2 heater 1 3 A heater current 0 3 A CO cell 5 mV after warm up and with fresh air 10 mV CO heater 7 5 V depending on flange temperature 3 0 V 1 V CO heater 0 4 A heater c...

Страница 21: ...user level 1 PIN code Changing the pin code is described in following chapter 9 6 1 PIN code level 1 Entering the factory default PIN code 1111 user level 1 gains access to the SET main menu and the...

Страница 22: ...ove cursor to values which can be changed EDIT change value by means of UP DOWN and POS OK save values BACK return to main menu Integration times from 0 to 30 seconds are settable where 0 seconds mean...

Страница 23: ...must always be the zero calibration with AIR The delivery state of this function is off 9 6 6 Set calibration cylinder values This function allows setting the values of span gas cylinders used for cal...

Страница 24: ...oid valves is possible By means of the key F1 the solenoid valve SV1 is switched ON or OFF key F2 the solenoid valve SV2 is switched ON or OFF and with the key F3 the single solenoid valve SV3 is swit...

Страница 25: ...96 4 20 mA range CO 1 000 ppm O2 cylinder1 21 00 O2 cylinder2 2 00 CO cylinder1 0 ppm CO cylinder2 1 000 ppm mixture gas with additional 50 ppm H2 gas is expected Pulse duration 2 sec Numb of pulses 6...

Страница 26: ...n the stack On site calibration is done without removing the probe from the stack After calibration tighten the calibration inlet port plug with a 14 mm wrench for 1 6 of a turn Perform adjustment 1 P...

Страница 27: ...scription 1 Electrical connector 2 Hose connection cal gas 3 Probe tube stainless steel 4 Flange 4 ANSI 150 or DN65PN6 5 Probe head transmitter 6 Calibration gas cylinder 7 Pressure regulator with flo...

Страница 28: ...al gas and back purge air 3 Probe tube stainless or Inconel steel 4 Flange DN65 PN6 or 4 ANSI 150 5 Probe head 6 Calibration gas cylinder 7 Pressure regulator with flow volume control 8 Transmitter 9...

Страница 29: ...ion cal gas and back purge air 3 Probe tube Inconel or ceramic 4 Flange 4 AINSI 150 5 Probe head 6 Calibration gas cylinder 7 Pressure regulator with flow volume control 8 Transmitter 9 Special connec...

Страница 30: ...ration adjustments on the OMS420 9 7 4 Pin Code level 2 Calibration with AIR zero gas or GAS span gas requires the use of level 2 pin code see chapter 9 6 1 POS position UP DOWN to change the numbers...

Страница 31: ...should be followed by new calibration using calibration gas cylinder Only authorized persons or manufacturer s trained technicians are authorized to perform adjustments on the OMS420 9 7 6 CO cell set...

Страница 32: ...l see picture below and select 10mV for offset trim and 100mV for span trim If actual values displayed are other than 10mV to 100mV setting can be adjusted by selecting Set offset or Set span and edit...

Страница 33: ...hange value by means of UP DOWN until you read 400mV up to2000mV on the multi meter OK save values BACK return to calibration menu Note The 4 20mA analog output factory adjustment can deteriorate over...

Страница 34: ...bubblier With single component gas CO the calibration is not possible If no mixture gas is available then see chapter 9 7 11 Calibration of O2 only set CO cylinder value 0 no calibration Calibration...

Страница 35: ...ill be displayed Read calibration gas cylinder values and if any change is required other cylinder values then edit O2 and CO in the display by pressing O2 then the CO key Wait approximately 5 minutes...

Страница 36: ...nstrument will display same value as the comparing CO selec tive measuring instrument 1 The measured value by means of Delta65 can be set desired value by pressing EDIT key 2 Using the UP and DOWN key...

Страница 37: ...or regulation factory only PREV or NEXT move cursor EDIT change value by means of UP DOWN and POS OK save values BACK return to calibration menu First line regulation point actual CO Second line new s...

Страница 38: ...ay Configuration not ok Checksum wrong New adjustment required Display Flash error New adjustment required Please contact your dealer manufacturer Display Unit to warm Hot environment Ambient temperat...

Страница 39: ...2 0 2 O2 or 5 of reading whichever is larger COe not selective measurement accuracy is not an issue Response time 10 seconds Electronics micro controlled based on board graphic backlit LCD Output sign...

Страница 40: ...0 1A CO sensor not o k 64 Heating current 0 1A CO sensor not o k 128 Calibration error calibration factor cannot be set Test gas cylinder is empty filter dirty or clogged Calibration offset value cann...

Страница 41: ...ting point value reads 1E38 when not available Defined registers Available data with modbus command 4 READ INPUT REGISTERS PLC address Protocol address Data type Number of registers Register content D...

Страница 42: ...OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420 44 12 3 Declaration of confirmity www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 43: ...OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420 45 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Отзывы: