OPERATING MANUAL OXYGEN MONITORING SYSTEM
OMS 420
24
9.6.2
Set analog output
PREV or NEXT
move cursor to values which can be changed
EDIT
change value by means of UP/DOWN and POS
OK
save
values
BACK
return to main menu
Factory defaults are:
20mA: O
2
0 – 21.00 %
20mA: CO
0 – 1.000 ppm
Hold after purge
typical 20 sec. (from 5 to max 200 seconds)
Notice:
The last measured value on analog output 4-20mA after back-purge is hold even after back-purge cycle is finished.
9.6.3
Damping
This function allows user to set a damping (averaging the measurement over a period of time) of measurement, in the display as
well in the analog output.
PREV or NEXT
move cursor to values which can be changed
EDIT
change value by means of UP/DOWN and POS
OK
save
values
BACK
return to main menu
Integration times from 0 to 30 seconds are settable, where 0 seconds means no damping.
9.6.4
Back-purge setup
This function allows setup of the optional back-purge solenoid valve to control:
•
interval time between back purges (hours and minutes)
•
pulse duration or interval time for energizing the solenoid valve (opening the valve)
•
number of pulses during one back purge cycle
PREV or NEXT
move cursor to values which can be changed
EDIT
change value by means of UP/DOWN and POS
OK
save
values
BACK
return to main menu
If hours and minutes are 0, the back purge function is deactivated. The delivery state of this function is off.