3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
Section 5
Maintenance
Item
Inspection
Interval
Daily
Weekly
Connection status of power cord or plug
•
Damages in power cord or plug
•
Cleanliness of top plate
•
Cleanliness of main body and accessories
•
Damages in plate, switches, buttons, Controller
•
Heating capability check (up to 500
℃
)
•
Stirring capability check (up to 1200 rpm)
•
Assembly status of all parts or accessories
•
Cleaning Product
WARNING
•
Never immerse this unit in water or any other liquid.
•
Do not allow any liquid or wet material to get inside the unit
when cleaning.
•
Do not reconnect this unit to power outlets until all cleaned
surfaces have dried.
CAUTION
•
Do not use chlorine bleach, ammonia-based cleaners,
abrasives, ammonia, or metal scouring pads when cleaning.
•
During cleaning and general operation take care not to scratch
the surface of the ceramic-coated top plate as this could result
in subsequent thermal breakage.
•
The ceramic top may break during use if not properly
maintained.
•
Keep the ceramic top clean. A nonabrasive cleaner may be
used to clean the ceramic top
•
Inspect the ceramic top for damage during cleaning.