5-3 Manual Datalogger (Set sampling time = 0
second)
a. Set sampling time is to 0 second
Press the " LOG Button ( 3-8, Fig. 1 )>1.5 second, the LCD
will show the indecator " LOGGER " and " Position no. " symbol
then press the " LOG Button " ( 3-8, Fig. 1 ) once , the
" LOGGER " symbol will flashing once and Beeper will sound once
at the same time the measuring data along the time
information will be saved into the memory circuit .
Remark :
During execute the Manual Datalogger, it can use the "
▲
Button " ( 3-5, Fig. 1) or "
▼
Button " ( 3-6, Fig. 1 ) to
set the measuring position ( 1 to 99, for example room 1
to room 99 ) to identify the measurement location , the
Top Display will show P x ( x = 1 to 99 ).
b. Finish the Datalogger
During execute the Datalogger function, press the " LOG
Button ( 3-8, Fig. 1 ) > 1.5 seconds continuously again will
finish the Datalooger function, the Position no. " PXX " will
be disappeared and finish the Datalogger function.
5-4 Check time information
During the normal measurement ( not execute the
Datalogger ), If press " Time " ( 3-7, Fig. 1 )
once , the LCD display will present the time
information of Year/Month/Date ,Hour/ Minute/Second
and the Sampling time information in sequence.
11
Содержание MG-3003SD
Страница 16: ...EXCEL graphic screen for example 14...