4-3 Checking Speed (RPM/deg/ms) ---External signal MODE
1) Connect signal wires with
main body of the instrument
after soldering the wires with
connectors.
Fig.2
2) Firmly plug power cord into a 110VAC (USA) or
220VAC (Europe) single phase outlet.
3) Turn power switch on
4) use int/Ext. button. " ( 3-3, Fig. 1 )" select to External Signal
MODE. And use " MODE button ( 3-6, Fig. 1 )" select to as
following application function.
RPM Mode
When receiving the input signal, the flashing will start and the input
signal will be converted and displayed in rpm. At this moment the
Digital adjusting knob does not interfere
degree , ms Mode ( Delay Mode )
When the input signal cycle is 360°, the flashing can be delayed
by 1~359°.
The delayed angle is adjusted by the Digital adjusting knob.The display
unit can be selected in degree or ms. LED indecator will light up.
deg flashing --shows angle display
ms flashing -- shows time, converted from angle
11