Operation Manual
7 of 18
The sensor under test must always be removed from the calibrator after use.
There is risk that the sensor may become stuck.
Component life time can be shortened by continuous high temperature
operation.
Use of this instrument at HIGH TEMPERATURES for extended periods of time
requires caution.
Do not turn off the instrument at temperatures higher than 100˚C. This would
create a hazardous situation. Select a set-point less than 100˚C and allow the
Instrument to cool before turning it OFF.
Check that earth connection for the instrument is present and attach Power
Cord.
Don’t exceed the pressure supplied to Forced Air Cooling System by 7 Kg/cm
2
D.
HOT SURFACE:
Do not touch the well access surface of the instrument.
The block vent may be very hot due to the fan blowing across the heater block of
the instrument.
Do not touch the Well or the Insert while the calibrator is heating up, they may
be very hot.
Do not touch the tip of the sensor when it is removed from the Insert / well, it
may be very hot.
Don’t switch off the calibrator immediately after use. You must wait until the
instrument reaches a temp at 100˚C before you switch off.
The calibrator is designed to offer a minimum of basic insulation only.
Recommended usage of the calibrator – 8 hours a day.
When there is power failure during calibration, the calibration can be started from
the reading from where the power is cut off.