![MRC BWLC-A1-C2 Скачать руководство пользователя страница 72](http://html.mh-extra.com/html/mrc/bwlc-a1-c2/bwlc-a1-c2_instruction-manual_1845694072.webp)
SU_A CR LF
SU_E CR LF
- command accepted and in progress
- timeout while waiting for stable results
SU_I CR LF
- command comprehended but cannot be executed
SU_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in current unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
space
stability
space
sign
mass
space
unit
CR
LF
Example:
S U CR LF
– computer command
S U _ A CR LF
- command accepted and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
- command done, mass
value in current unit is returned.
20.4.7. Send the result immediately in current unit
Syntax:
SUI CR LF
Possible answers:
SUI_I CR LF
- command comprehended but cannot be executed
MASS FRAME
- mass value in current unit is returned immediately
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
Example:
S U I CR LF
– computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command executed
and mass returned
- 72 -
Содержание BWLC-A1-C2
Страница 2: ...www mrclab com 2 ...
Страница 3: ...MARCH 2010 3 ...
Страница 11: ...Scales BWLC C1 R Scales BWLC C2 R Scales BWLC C1 K Scales BWLC C2 K 11 ...
Страница 31: ...Return to weighing see 11 2 2 14 3 Baud rate Procedure Enter the submenu P2 Prnt and then 31 ...
Страница 46: ...but deviation of load mass placed on the pan referring to the mass of standard in 46 ...