- Zero value from the AD converter
FULL2
- Measurement range
overflow
LH
- Start mass error, the mass on the weighing platform is
beyond the acceptable range (-5% to +15% of start mass)
5-FULL
- Display range overflow in totalizing
Notice:
1.
Errors:
Err2
,
Err3
,
Err4
,
Err5
,
Err8
,
null
, that appear on the display
are also signalled by a short beep sound (about 1 sec.);
2.
Error
FULL2
that appears on the display is also signalled by
a continuous sound until the cause of error disappears.
22. TROUBLE SHOOTING
Problem
Cause
Solution
Turning on does not
work
Discharged batteries.
Connect to mains or
change batteries
No batteries (not installed or
improperly installed)
Check the correctness of
installation (polarization)
The scale turns off
automatically
„
t1” set to „YES” (Power
save)
In „othr” submenu change
„5.4 t1” to „no”
After turning on „LH”
message on the display
Loaded weight pan during
powering up
Unload the pan. Then the
scale will indicator zero.
- 77 -