T_A CR LF
T_D CR LF
- command accepted and in progress
- command completed
T_A CR LF
T_v CR LF
- command accepted and in progress
- command comprehended but tare range overflow appeared
T_A CR LF
T_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
T_I CR LF
- command comprehended but cannot be executed
20.4.3. Get tare value
Syntax:
TO CR LF
Possible answers:
TO_TARA
CR LF
- command executed
Frame format:
1
2
3
4
5-6
7-15
16
17
18
19
20
21
T
O
space
stability
space
tare
space
unit
CR
LF
Tare
- 9 characters with decimal point justified to the right
Unit
- 3 characters justified to the left
20.4.4. Send the stable result in basic unit
Syntax:
S CR LF
Possible answers:
S_A CR LF
S_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in basic unit is returned
- 70 -