reduce  the sampling  process to reduce  the airway residue. Please  click "

"  "

"  key to 

change the self-purification time, and then press "Start" to start the self-purification. Note 

that during the self-purification process, the barometric pressure source is still an external 

compressed gas (tank). Do not disconnect the external compressed gas. In addition, if you 

still  need to continue  to repeat  the same  compressed gas sampling, you can not directly 

press "back" to start the next sampling. 

(11)

 

Sampling  is completed: Please  close the external compressed gas (tank) after the

 

valve, then the intake duct from the back of the instrument pull out. 

Sampling precautions.3

 The 

AIS-4

 of compressed air microbial samplers can not be used for toxic,

 

corrosive,

 

(1)

 flammable and explosive gases, if necessary, please contact the manufacturer or agent to

 . purchase other models

Sterilization: The porous sampling head can be sterilized by autoclaving at 121 

°

C

 

(2)

for 30 minutes (or wiped with 75% alcohol) at the high pressure steam sterilizer. The gas 

dispersing  head, the  Petri dish and the appearance  of  the  instrument are  also used 75% 

alcohol for wipe sterilization. After each sampling test, if you want to continue sampling 

another compressed gas (cans), you need to repeat the sterilization treatment. 

(3)

 

When installing the Petri dish, the dish must be stuck in the center of the base and the

 

bottom of the dish is pressed down to the base. 

(4)

 

Pressing  the  "return"  key  before  sampling is complete will  forcibly  terminate the

 

sampling  operation. The  internal airway solenoid  valve will  be  automatically closed and 

the sampling volume will be stopped, but the sampling record will be saved.   

4.Device instructions

Catheter Interface: Gas duct interface  with easy to plug  can be  replaced design, pull  out 

when you need to press the interface of the blue plastic ring, and then pull out the catheter, 

and insert the  catheter  directly  into the reverse  pull  the catheter To  strengthen  the blue 

plastic ring sealing effect. 

Pneumatic control valve: the left side of  the  instrument for  the  air pressure  adjustment 

Содержание AIS-4

Страница 1: ...Compressed Air Sampler Instruction ISO9001 2008 Certified Enterprise PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Страница 2: ...oper use of this equipment please keep it for future reference Directory I Equipment Introduction 1 II Application 1 III Working Environment 1 IV Main Features 1 V Technical Parameters 2 VI Product Li...

Страница 3: ...etsec 23 1 to 23 5 m sec at 200 L min The surface of the medium is equivalent an 1hAnderson Principle 5 Grade 6 at 200L min ensuring that all particles larger t ish wasdmicron 0 65 micron at 200L min...

Страница 4: ...pling design the sampling head with up to 300 microporous so that micro organisms evenly distributed in the surface of the medium to reduce the dust bacteria overlap effectively reduce the microbial c...

Страница 5: ...set Lithium battery charger DC 8 4V 1000mA 1set Manual certificate warranty card 1 set USB data cable 1 Piece VII Structural Features 1 The composition of the sampler includes gas dispersing head sam...

Страница 6: ...4 quick connector gas dispersing head sampling head air pressure regulating valve control panel gas flow meter outlet air inlet switch charging interface purifier internal...

Страница 7: ...ired for microbial culture according to the requirements of the experiment and prepare the corresponding plate Before starting sampling make sure that the storage temperature of the sampling petri dis...

Страница 8: ...state please open the external compressed gas tank valve compressed gas will be through the air inlet pressure control valve gas flow See section 4 of this chapter After adjustment press Start to proc...

Страница 9: ...or 30 minutes or wiped with 75 alcohol at the high pressure steam sterilizer The gas dispersing head the Petri dish and the appearance of the instrument are also used 75 alcohol for wipe sterilization...

Страница 10: ...lly lock the baffle screws 5 the system function menu description Working mode selection In the main interface press key to select the working mode as program control manual control self purification...

Страница 11: ...all sampling records Please choose carefully View the record In the main interface press the Select button to enter the record query interface you can view all the sampling record data including sampl...

Страница 12: ...thod not mentioned above consult the manufacturer or agent to determine if the method used will damage the instrument 6 The display shows that when the battery power is low the battery is displayed as...