knob, adjust the need to pull the knob out slightly, counterclockwise rotation to open the 

air flow channel, clockwise rotation to close the air flow channel. 

Gas  flowmeter:  the  right  side  of  the  gas  flow  meter  below  a  flow  adjustment  knob,  the 

same counter-clockwise rotation to open the air flow channel, clockwise rotation to close 

the air flow channel. 

Purifier: the purifier for the consumables, after a long period of use, the purification effect 

will  be  reduced,  it  is  recommended  to  use  the  time  accumulated  for  200  hours  or  more 

than a year after the replacement. When replacing, remove the top side of the baffle screws, 

remove the purifier from the instrument, remove the clamp with a screwdriver to remove 

the  purifier  tube,  take  a  new  purifier  and  push  it  back  into  the  instrument.  Into  the 

compressed  gas  into  the  self-purification  mode  to  check  the  phenomenon  of  no  leakage, 

and finally lock the baffle screws. 

5.the system function menu description

Working mode selection 

In  the  main  interface,  press  "

"  "

"  key  to  select  the  working  mode  as  "program 

control", "manual control", "self-purification mode" 

Programmed Sampling: The recommended program controls the sampling mode. For 

details, see Section 2 of this chapter. 

Manual  control:  to  meet  some  of  the  user's  manual  control  of  special  needs,  in  this 

mode, press the "start", enter the manual control interface, the bottom of the screen shows 

the  current  two  airway  solenoid  valve  status,  press  "

"  Change  the  state  of  the  airway 

solenoid valve, press the "Select" button to switch to select another airway solenoid valve. 

Self-purification mode: Press "

" "

" key to change the self-purification time, press 

the  "Start"  button  to  start  self-purification.  In  the  access  to  compressed  gas  (cans) 

conditions,  open  the  self-purification  after  the  gas  through  the  air  inlet 

  purifier 

 

pressure control valve 

  gas flow meter 

  sampling head, through the air flow to the 

sampling  process  of  exhaust  airway  residue.  (Equivalent  to  the  "open  air  open"  state  in 

manual control mode). 

Содержание AIS-4

Страница 1: ...Compressed Air Sampler Instruction ISO9001 2008 Certified Enterprise PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Страница 2: ...oper use of this equipment please keep it for future reference Directory I Equipment Introduction 1 II Application 1 III Working Environment 1 IV Main Features 1 V Technical Parameters 2 VI Product Li...

Страница 3: ...etsec 23 1 to 23 5 m sec at 200 L min The surface of the medium is equivalent an 1hAnderson Principle 5 Grade 6 at 200L min ensuring that all particles larger t ish wasdmicron 0 65 micron at 200L min...

Страница 4: ...pling design the sampling head with up to 300 microporous so that micro organisms evenly distributed in the surface of the medium to reduce the dust bacteria overlap effectively reduce the microbial c...

Страница 5: ...set Lithium battery charger DC 8 4V 1000mA 1set Manual certificate warranty card 1 set USB data cable 1 Piece VII Structural Features 1 The composition of the sampler includes gas dispersing head sam...

Страница 6: ...4 quick connector gas dispersing head sampling head air pressure regulating valve control panel gas flow meter outlet air inlet switch charging interface purifier internal...

Страница 7: ...ired for microbial culture according to the requirements of the experiment and prepare the corresponding plate Before starting sampling make sure that the storage temperature of the sampling petri dis...

Страница 8: ...state please open the external compressed gas tank valve compressed gas will be through the air inlet pressure control valve gas flow See section 4 of this chapter After adjustment press Start to proc...

Страница 9: ...or 30 minutes or wiped with 75 alcohol at the high pressure steam sterilizer The gas dispersing head the Petri dish and the appearance of the instrument are also used 75 alcohol for wipe sterilization...

Страница 10: ...lly lock the baffle screws 5 the system function menu description Working mode selection In the main interface press key to select the working mode as program control manual control self purification...

Страница 11: ...all sampling records Please choose carefully View the record In the main interface press the Select button to enter the record query interface you can view all the sampling record data including sampl...

Страница 12: ...thod not mentioned above consult the manufacturer or agent to determine if the method used will damage the instrument 6 The display shows that when the battery power is low the battery is displayed as...

Отзывы:

Похожие инструкции для AIS-4