IMPORTANT SAFETY PRECAUTIONS - continued

WARNING! 

For your safety, the information in this manual must be followed to 

minimize the risk of fire, electric shock, or to prevent property damage, personal injury, or 
loss of life. 

WARNING! 

Make sure the bed is in the lowest position when the patient is left 

unattended. This can reduce the severity of any injuries resulting from patient falls.

WARNING! 

Except when the bed is being moved, the brakes should always be set 

(locked). After applying the brakes test the motion of the bed to ensure that the brakes are 
properly engaged. Patients use the bed for support when getting in or out of bed, and could 
be injured if the bed unexpectedly moves sideways. 

WARNING! POSSIBLE INJURY OR DEATH. 

This bed is intended for 

use as an adjustable mattress deck that is designed for patient and caregiver comfort and 
convenience. Use of this product in any way that it was not designed for could result in injury 
or death.

WARNING! 

Patients may become entangled in the hand pendant cord. Patients with 

diminished mental acuity should not be allowed access to the pendant. Failure to do so could 
result in injury or death.

WARNING! 

The safe working load of this bed is 500 lbs evenly distributed, which 

includes the weight of the patient, mattress and all accessories and visitor. Also jumping, 
falling on or standing on the bed will concentrate the load and might result in property 
damage, personal injury or death. 

WARNING! 

Ensure that no arms, hands, legs or feet are under when lowering the 

mattress frame decks.  

WARNING! POSSIBLE INJURY OR DEATH 

Use only the proper size 

mattress in order to minimize the gap between the side of the mattress and the assist device. 
The gap must be small enough to prevent the patient from getting his/her neck caught in 
this location (read Entrapment Guidelines for further information).

Cust
omer 

Supp

or

t

M

ain
tenanc

e

O

pe

ra

tin

g I

ns

tru

ctio

ns

Troublesho

oting

Sa

fet

y In

stru

ctio

ns

As

se

m

bly

WARNING! 

 Care must be made to ensure that the patient does not fall-on or slide-over the 

mattress  side stops which may cause injury.  

WARNING!

 

When lowering the bed, ensure that no person, limbs, bedding or any 

foreign objects are protruding between the bed frame and the floor. Ensure that Over bed 
tables and patient lifts are safely removed. 

Содержание QD7000 Series

Страница 1: ...cifications 6 Unpacking and Assembly 8 Operating Instructions 13 Maintenance 18 TroubleshootingTips 20 Warranty Claims 23 Parts Order Form 24 Replacement Parts Warranty 26 36 WARNING To avoid injury y...

Страница 2: ...tions and Icons Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly Icons and Symbols used on the Valet Bed DANGER CAUTION and WARNING symbol is used to id...

Страница 3: ...ental acuity should not be allowed access to the pendant Failure to do so could result in injury or death WARNING The safe working load of this bed is 500 lbs evenly distributed which includes the wei...

Страница 4: ...a spill immediately Unplug the bed from the power source and remove it from service Remove the patient from the bed and clean the spill Remove the bed from service until the bed is tested and determin...

Страница 5: ...bed system where there is a potential for patient entrapment Entrapment may occur in flat or articulated bed positions with the rails fully raised or in intermediate positions The seven entrapment zo...

Страница 6: ...length 81 inches including bumpers and Head Foot Boards 2 Sleep surface 74 inches BedWidth Overall width side rails raised 41 1 8 inches Mattress Deck 1 Width 35 inches 2 Length 74 or 80 inches Mattr...

Страница 7: ...Head EndWall Stand Off Bumpers High impact absorption bumpers Swivel Casters Heavy duty 3 2 total locking and two swivel casters mobility at any bed height Optional Immobilizer Legs Built in stabilize...

Страница 8: ...ring in place care must be taken so that only the shipping ties are removed These ties can be identified by the fact that they have not been trimmed Remove cardboard sleeve protecting the foot end of...

Страница 9: ...xtender attaches directly to the head deck If a headboard is to be used with the extender then a set of headboard extender brackets are required to offset the headboard and allow clearance for the ext...

Страница 10: ...asier Step 2 Locate the proper holes in each corner of the bed some models use side stops as illustrated below Step 3 Place a Mattress Stop 1 in one corner overlapping the pre punched holes as illustr...

Страница 11: ...THANE ROLLER 2 4 QS 3560 ROLLER BUMPER BRACKET 2 5 QP 1748 COTTER RING RUE 1 2 2 6 1854 LOCKNUT 5 16 18 4 Procedure Step 1 Mount Roller bracket 4 on the head end of the bed frame as shown using two he...

Страница 12: ...attach the connector to the pins visible in the end channel below Take care to get the connectors lined up properly 1 2 3 4 5 6 7 1 2 Foot End Control Setup www mchealthcare com Attach the bolt 1 and...

Страница 13: ...d Length Siderails Lifting the siderail up completely will latch it into the upright position To lower the rail you must press the release lever which is located at the foot end of the lower channel b...

Страница 14: ...atients tend to use the bed for support when getting in or out of bed Patients could be injured if the bed moves sideways to guard against this risk you must engage the Directional Lock Hinges Simply...

Страница 15: ...gently raise the patient s Knees and Feet press and hold the Foot UP button on the pendant To gently lower the patient s Knees and Feet press and hold the Foot DOWN button on the pendant Auto Contour...

Страница 16: ...e functions from the foot end ensure that the function is not locked out the red light should be off and then press the function up or down as required When the keypad is disconnected then all functio...

Страница 17: ...battery port on the controller There is a retaining clip securing the ports which must be removed prior to inserting the battery cable Patient Helper The patient helper consists of a bracket mounted t...

Страница 18: ...operation over full range The bed should move evenly through its full range of motion from low to high position and from flat to contoured with no jerking or twisting Grinding popping other noises The...

Страница 19: ...r e d i S s l e n a p d r a o b t o o F d a e H h s i n i f t n i a P e c n a r a e p p a l l a r e v O s l a i t i n I s t n e m m o C Customer Support Maintenance Operating Instructions Troubleshoot...

Страница 20: ...Hi Lo White 6 Hi Lo motor Blue 7 AC power inlet 1 If a nurse lock out box is present ensure that it is unlocked If it is not then turn the function switch to the unlock position and the function shou...

Страница 21: ...pull to remove the clip Ensure that all of the connectors are firmly seated and then check to see if this corrects the problem Try unplugging the function cable port 2 in diagram and using a hand pend...

Страница 22: ...use port by pulling on the tab straight up with a pair of pliers and replace both fuses using a pair of long nose pliers or a fuse replacement tool Spare fuses are included in the fuse cover and repla...

Страница 23: ...only accept cleaned parts 2 3 4 5 6 7 8 9 M C Healthcare Products will receive orders for warranty claimed products M C will assign an RMA number for parts that are to be returned to support the warr...

Страница 24: ...24 Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly NOTES...

Страница 25: ...ame Bed Serial Number s Facility Name Address Phone Fax For an RMA number call M C Healthcare Customer Service at 1 800 268 8671 Fax 1 905 563 8680 or E mail beds mchealthcare com the completed Form t...

Страница 26: ...26 Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly REPLACEMENT PARTS www mchealthcare com 1 MAIN FRAME COMPONENTS 2 FRAME MOTOR ASSEMBLY...

Страница 27: ...port Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly 27 REPLACEMENT PARTS 3 FOOT END RAISING MECHANISM QP1482 15 QP1305 Nylon ShoulderWasher 2 16 1826 3 8 16 UNC Nylock...

Страница 28: ...28 Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly REPLACEMENT PARTS www mchealthcare com 4 VASCULAR FOOTREST ASSEMBLY 5 CASTER BRACKET ASSEMBLY 4011...

Страница 29: ...Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly 29 REPLACEMENT PARTS 6 HEAD END RAISING MECHANISM...

Страница 30: ...30 Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly REPLACEMENT PARTS www mchealthcare com 7 MATTRESS DECKS QP1996 QS3452...

Страница 31: ...Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly 31 REPLACEMENT PARTS 8 HALF LENGTH SIDE RAIL 9 ASSIST RAIL...

Страница 32: ...32 Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly REPLACEMENT PARTS www mchealthcare com 10 HEAD BOARD 11 FOOT BOARD...

Страница 33: ...Customer Support Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly 33 REPLACEMENT PARTS 11 FOOT BOARD PENDANT HOLDER 12 SIDE RAIL MOUNTED PENDANT HOLDER QP1561...

Страница 34: ...pport Maintenance Operating Instructions Troubleshooting Safety Instructions Assembly REPLACEMENT PARTS www mchealthcare com 13 SIDE RAIL MOUNTED GOOSENECK PENDANT HOLDER 14 HEAD FOOTBOARD MOUNTED GOO...

Страница 35: ...NSERT 2 2 QP 1842 3 TOTAL LOCK CASTER 3 2 QP 1841 W 150 FLAT WASHER 4 2 QP 1847 HINGE ASSEMBLY 5 2 QP 1846 1 1 4 ROUND CAP PLUG 6 2 QS 3306 VALET LEG TUBE 5 780 7 2 2004 BLACK PLUG 8 1 QS 3486 MOBILIT...

Страница 36: ...for removing or reinstalling parts which are the purchaser s expense This warranty does not apply where The product has been abused misused or improperly maintained The product has been exposed to exc...

Отзывы: