The Valet bed can be cleaned with warm water and a mild detergent or diluted disinfectant. Use a damp
cloth or a sponge to clean all exposed surfaces of the bed. Take care not to use excessive water around
the control panels and the control box.
The Valet bed was not designed to be cleaned using a high pressure wash tunnel.
Preventive Maintenance
There are a number of maintenance items that should be reviewed every 6 months to ensure proper
functioning of the bed.
LUBRICATION – All pivot points and bolted rotating joints should be lightly oiled with a penetrating
lubricant containing an extreme pressure additive (e.g. Castle Iron Man). A lithium based grease is not a
suitable alternative for this application. No lubrication is required on the motor itself, but the clevis pins
may be lubricated if required. Remove any excess oil/grease after application with a cloth.
MOTOR OPERATION– The motors should run smoothly and quietly throughout their range.
Normal operation over full range - The bed should move evenly through its full range of motion from
low to high position and from flat to contoured with no jerking or twisting.
GRINDING, POPPING, OTHER NOISES – The bed should be fairly silent. Some motor/gear noise is
normal, and some settling noises are expected, but nothing loud and no squeaking. Look for abnormal
wear, lubricate any joints exhibiting noise.
BENT PARTS – Check for abnormally bent parts.
RETAINING RINGS – All clevis pins should have retaining rings present. Replace any missing pins.
FASTENERS – Ensure that all fasteners are tightened and seated completely. Do not
over-tighten the fasteners at the pivot points of the decks.
CABLING– Inspect the cabling on the bed to ensure that it is all still in place with no dangling or loose
wires, and that it is not interfering with any moving parts. Check for any cut, nicked or crushed wiring.
Reattach any loose wires, and repair or replace any damaged wires.
POWER CORD – Inspect the power cord for nicks or crushed insulation. Replace if damaged.
CASTER LOCKING – Check to ensure that the casters swivel and roll freely when unlocked, and do not
swivel or roll when the lock is engaged. Tighten, repair or replace if damaged, loose or not working
properly.
MAINTENANCE