mSquad User’s Guide 

 

 
 
 

2.

 

Battery 

Fully charge the POLI batteries upon receiving the mSquad and before each day’s use.    The Li-
ion batteries are charged using a Micro-USB cable.   
 

 

WARNING 

To reduce the risk of ignition of hazardous atmospheres, recharge, remove or replace the battery 
only in an area known to be non-hazardous! 

2.1 Battery Charging 

 

Up to four POLI monitors can be charged simultaneously using the cables inside the mSquad 
carry case. Plug the mini ends of the Micro-USB cables into the charging port of each POLI, and 
the power adapter into a wall socket (100-240 VAC). The POLIs can also be charged 
individually on a personal computer using the USB power/communications cables provided. 

NOTE

: Any locally-obtained USB A to Micro B USB cable works for charging, but does not 

work for communication with mPower Suite configuration and data transfer software. The 
mPower USB cable P/N M-011-3003-000 is required for a PC to recognize the instrument and 
communicate with mPower Suite. 
 
 
 
 
 
 
 
 
 
 
 
The screen will display a battery icon from empty to full and one alarm LED will remain yellow. 
When the battery is fully charged, the icon displays full grid status (see Section 3.1), and the 
alarm LED turns green. 
 

2.2 Battery Status   

The battery icon on the display shows the battery level, and alerts of any charging problem. 
 

 

 

 

 

blink 

Full charge

 

2/3 charge

 

1/3 charge

   

Battery Low

 

Battery Alarm

 

 
When the battery’s charge falls below a preset voltage, the instrument warns by beeping once 
and flashing once every minute. The instrument automatically powers down within 10 minutes, 
after which the battery must be recharged. When a low-battery alarm occurs, we recommended 
promptly switching instruments to a fully charged POLI, and/or charging the battery in a non-
hazardous location. 
 

Содержание mSQUAD MP400H

Страница 1: ...mSQUAD Mobile Wireless Multi Gas Team Warning System MP400S MP400H User s Guide Rev 1 0 January 2020...

Страница 2: ...t up 10 4 2 Alarm Notifications 10 4 3 Head Host Calibration 12 5 Configuration Mode 13 5 1 Navigating Configuration Mode to Edit Parameters 13 5 2 Man Down Function 14 5 3 Registering Soldier MP400S...

Страница 3: ...RRATIC READING MAY INDICATE A GAS CONCENTRATION BEYOND UPPER SCALE LIMIT WHICH MAY BE HAZARDOUS ATTENTION DES LECTURES SUP RIEURES A L CHELLE PEUVENT INDIQUER DES CONCENTRATIONS EXPLOSIVES TOUTE LECTU...

Страница 4: ...ions for Safe Use The POLI multi gas detector must be calibrated if it does not pass a bump test when a new sensor has been installed or at least once every 180 days depending on use and sensor exposu...

Страница 5: ...by diffusion because the communications antenna is attached to the pump inlet port while MP400S Soldier units have a built in sampling pump for active gas sampling and thus a somewhat faster response...

Страница 6: ...work for communication with mPower Suite configuration and data transfer software The mPower USB cable P N M 011 3003 000 is required for a PC to recognize the instrument and communicate with mPower...

Страница 7: ...ration alarms 3 1 Turning On Off To turn on press and hold the Key for 3 seconds until the buzzer beeps and the red LED turns on To turn off in normal reading mode press and hold the key for a 5 secon...

Страница 8: ...ength at 0 5 level Pump status pump versions only Datalogging enabled cannot turn off Battery voltage status Man down alarm enabled All sensors have been bump tested and calibrated and none are overdu...

Страница 9: ...act as soft keys mapped to two text or symbol boxes at the LCD bottom that change numbers and make selections under various menus In addition to the functions described above the Left OK key can be u...

Страница 10: ...est of connectivity can be done by sending a Panic Alarm Hold Left key for 3 seconds 4 2 Alarm Notifications The POLI provides alarms including audible buzzer visible bright LEDs vibration and text no...

Страница 11: ...e alarm is silenced If a Soldier is turned off other Soldiers alarm immediately for 15 seconds The Host Head is alarmed with a symbol displayed next to the user name after the communication lost timer...

Страница 12: ...ut the user is OK it can be stopped using the Left key Clear WARNING The Man Down feature cannot detect a worker in distress in all situations even if they have collapsed The Man Down function should...

Страница 13: ...menus and OK Enter to enter a menu to edit the parameters in its submenu 5 1 1 Configuration Mode Menus and Sub menus Calibration Measurement Alarm Datalog Monitor Setup Wireless Exit Fresh Air Calib...

Страница 14: ...ng the POLI to a horizontal position or pressing the Left OK key The Man Down icon on the main screen verifies that the Man Down function is enabled 5 2 2 Man Down Warn Time This menu allows adjustmen...

Страница 15: ...e will appear on the Host Head Registered Device List with a next to the name The Host assigns default user names in the order they are registered e g USR00 USR01 USR02 To enter a personalized user na...

Страница 16: ...vice Ping Assign Worker 1 05496868 JACK Ping Worker Assign Worker 1 05496868 JACK JACK Worker Assign Assign Worker 1 05496868 JACK BILL Worker Assign Worker Pings Monitor to Verify Unit Assign Worker...

Страница 17: ...network 5 6 Channel Select the operating Channel from CH1 to CH16 Ensure that the same selection is made on the Head Host as on all Soldiers in the mSquad 5 7 mPlatoon Wireless Networks Future system...

Страница 18: ...Edit the configuration parameters as desired including those under the tabs for each sensor in the upper right Click Write to upload the configuration to the instrument Read allows downloading the cur...

Страница 19: ...ease of reading on the displays Click Write on the bottom of the Name panel to send the user names to the Host Head If the Write and Reset buttons are greyed out first click Read to enable them type...

Страница 20: ...d Troubleshooting For maintenance and troubleshooting please see the POLI User Guide 8 mSquad Technical Specifications All specifications and listed sensors are subject to change without notice Please...

Страница 21: ...mSquad User s Guide 21 Technical Support and mPower Contacts mPower Electronics Inc 3046 Scott Blvd Santa Clara CA 95054 Phone 408 320 1266 Fax 669 342 7077 info mpowerinc com www mpowerinc com...

Отзывы: