mSquad User’s Guide
2
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of
using, maintaining or servicing this product. The product will perform as designed only if it is
used, maintained and serviced in accordance with the manufacturer’s instructions. The user
should understand how to set the correct parameters and interpret the obtained results.
CAUTION
!
•
REMOVE MONITOR COVER ONLY IN AREA KNOWN TO BE NON-HAZARDOUS.
•
RECHARGE BATTERY ONLY IN AN AREA KNOWN TO BE NON- HAZARDOUS.
•
USE ONLY mPOWER’S RECHARGEABLE LITHIUM BATTERY P/N M004-3002-000.
•
USE OF NON-mPower COMPONENTS WILL VOID THE WARRANTY AND CAN
COMPROMISE THE SAFE PERFORMANCE OF THIS PRODUCT
•
SUBSTITUTION OF COMPONENTS MAY IMPACT INTRINSIC SAFETY.
CAUTION:
HIGH OFF-SCALE READINGS MAY INDICATE AN EXPLOSIVE
CONCENTRATION. ANY RAPID UP-SCALE READING FOLLOWED BY A DECLINING
OR ERRATIC READING MAY INDICATE A GAS CONCENTRATION BEYOND UPPER
SCALE LIMIT WHICH MAY BE HAZARDOUS.
ATTENTION:
DES LECTURES SUPÉRIEURES A L’ÉCHELLE PEUVENT INDIQUER
DES CONCENTRATIONS EXPLOSIVES.
TOUTE LECTURE RAPIDE ET POSITIVE,
SUIVIE D’UNE BAISSE SUBITE AU ERRATIQUE DE LA VALEUR, PEUT INDIQUER
UNE CONCENTRATION DE GAZ HORS GAMME DE DÉTECTION QUI PEUT ÊTRE
DANGEREUSE.
WARNINGS !
ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS INSTRUMENT HAS
BEEN ASSESSED FOR PERFORMANCE.
UNIQUMENT, LA PORTION POUR DÉTECTOR LES GAZ COMBUSTIBLES DE CET
INSTRUMENT A ÉTÉ ÉVALUÉE.
CAUTION:
BEFORE EACH DAY’S USE, THE SENSITIVITY OF THE COMBUSTIBLE
GAS SENSOR MUST BE TESTED ON A KNOWN CONCENTRATION OF METHANE
GAS EQUIVALENT TO 20 TO 50% OF FULLSCALE CONCENTRATION. ACCURACY
MUST BE WITHIN 0 AND +20% OF ACTUAL. ACCURACY MAY BE CORRECTED BY
A CALIBRATION PROCEDURE.
ATTENTION:
AVANT CHAQUE UTILISATION JOURNALIERE VERIFIER LA
SENSIBILITE AVEC UNE CONCENTRATION CONNUE DE METHANE EQUIVALENTE
A 20-50% DE LA PLEINE ECHELLE. LA PRECISION DOIT ETRE COMPRISE ENTRE
0-20% DE LA VALEUR VRAIE ET PEUT ETRE CORRIGEE PARUNE PROCEDURE
D’ETALONNAGE.