Appendix B
Messages
21
Value Function
Comment
0
NOT READY
Unit is powering-up, or there is some problem
1
READY
Ready to start a test
XII
2
TESTING
Test in progress
3
WAITING
Waiting between tests
XIII
128
FAULT OPTICAL
LED failure / sensor blocked / filled with air
129
FAULT FLOW LOW
Flow too low for reliable test
XIV
130
FAULT FLOW HIGH
131
FAULT LOGGING
Fault with data logging
132
FAULT WATER SENSOR
Water sensor failure
Table III
The
STATUS CODE
Parameter
Status Flags Bitmask
This represents the states of various items in a bitmask format.
Bit
Function
Comment
0
RESULT_VALID
Current result is valid
1
RESULT_NEW
A new result is available
2
RESULT_LOG
Current result should be logged
3
TESTING
Test in progress
4
COMPLETE
Test is complete
5
ALM_HI_COUNT
High particle count alarm
6
ALM_HI_H2O
High water content alarm
7
ALM_HI_TEMP
High Temperature alarm
8
ALM_LO_COUNT
Low count alarm
9
ALM_LO_H2O
Low water content alarm
10
ALM_LO_TEMP
Low temperature alarm
11
REMOTE_CONTROL
Unit is under remote control
12
IO_IP
Start signal input
13
IO_OP1
Alarm output 1
14
IO_OP2
Alarm output 2
15
UNUSED
Not Currently Used
Table IV
Status Flags
Bits 0-2 are so that external equipment (for example LPA-View or a PLC/MMI)
can display, update and log results intelligently.