C A L I B R A T I O N S E T - U P
Equipment and methods used during the calibration process of any transducers or analyzers
are critical and can dramatically affect the quality, accuracy and performance of your analyzer.
Maintain all your calibration equipment in top condition, and always perform the calibration
process in a controlled environment.
Key sequence to access the Calibration Set-Up:
Set-Up
(Mode + Unit)
Scroll Up
until you display CAL
Enter
to access the program
ACCESS CODE
The Access Code is “2003”, this can not be
changed. In order to prevent tempering or deletion
of calibration data, we recommend releasing the
code number to limited personnel.
Scroll Up
to change the number
Enter
to validate
CAPACITY UNIT
This unit of measurement corresponds to the
transducer to be calibrated.
Scroll Up
to change units
Enter
to validate
HIGH CAPACITY
This value is the transducer’s maximum capacity
provided by the manufacturer. Always start with
the left most digit.
Scroll Up
to update digit value
Scroll Right
to select digit to be updated
Enter
to validate
B
+
B
Torque
Manual
Clear
Set-up
B
Auto
Clea
Downlo
+
B
Torque
Manual
Clear
Set-up
B
B
Auto
Clea
Downlo
+
B
Manual
Clear
Set-up
Torque
B
Aut
o
Manual
Clear
Downl
Set-up
+
B
Torque
Threshold
Note:
Make sure the Cell number corresponds
to the transducer you are intending to
calibrate. See Cell Set-Up section if you
need to change number.