![Mountz TorqueLab Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/mountz/torquelab/torquelab_manual_1840935012.webp)
12
This function is primarily used for testing and
calibrating click type mechanical torque wrenches. The
TorqueLab
â
captures the point where the wrench
clicks. This peak may be used for operator training on
correct use of the wrench. Always apply torque
smoothly to avoid false first peak readings.
U N I T S E L E C T I O N
The TorqueLab
â
analyzer supports seven units of torque measurement, lbf.ft, lbf.in, ozf.in, N.m,
cN.m, Kgf.m, and kgf.cm. However certain torque values cannot be converted into other units
due to display resolution, in these cases the TorqueLab
â
will warn you by displaying “Err”.
Key sequence to change units:
•
Units
•
Scroll Up
•
Enter
Torque Conversion Chart
Units to
Convert
ozf.in
lbf.in
lbf.ft
mN.m
cN.m
N.m
gf.cm
kgf.cm
kgf.m
mN.m
0.142 0.009 0.0007
1
0.1
0.001
10.2
0.01 0.0001
cN.m
1.416 0.088 0.007
10
1
0.01
102
0.102 0.001
N.m
141.6 8.851 0.738 1000
100
1
10,197 10.2 0.102
ozf.in
1 0.0625
0.005 7.062 0.706 0.007 72 0.072
0.0007
lbf.in
16 1 0.083
113 11.3
0.113
1152.1
1.152
0.0115
lbf.ft
192 12 1 1356 135.6 1.356 13,826
13.83
0.138
gf.cm
0.014 0.0009 0.00007 0.098
0.01
0.0001
1
0.001 0.00001
kgf.cm
13.89
0.868
0.072 98.07 9.807 0.098 1000
1
0.01
kgf.m
1389 86.8 7.233 9807 980.7 9.807 100,000 100
1
Lbf.in
Downlo
+
B
Auto
Clear
Pulses
American S.I.