Replacing the Trigger Assembly
To replace the trigger assembly:
1.
Remove the finger strap assembly.
2.
Turn the RS419 upside-down.
3.
Remove the set screw with
screwdriver.
4.
Turn the trigger assembly
counter-clockwise until the exit
window and scan trigger align.
5.
Lift the trigger assembly off the scan
assembly.
6.
Align replacement trigger assembly
with scan assembly.
7.
Rotate trigger assembly 1/4 turn
clockwise.
8.
Tighten the set screw with
screwdriver.
9.
Replace finger strap assembly.