background image

 

28

10.2 Prevention against water and moisture 

The microscope must be kept away from sources of water including water faucets, tanks, sinks, etc. 

The instrument should be kept in a low humidity environment (relative humidity should be kept below 

85%). If the eyepieces and other optical components are not to be used for an extended period of time, 

they should be placed in a dry cardboard box with the addition of a desiccating agent. If the 

environmental conditions of the current location are exceedingly moist and humid, Motic recommends 

use of a dehumidifier within the lab. 

 

10.3 Proper cleaning techniques 

 

If the surface of an optical lens is dusty, use a pipette to blow off the dust. 

 

If fingerprints or oil stains are discovered, or the use of a pipette is unable to remove dirt, wipe with  

 

an oil-free cotton swab or lens cloth lightly dipped in a mixture of pure alcohol and ether (mixed  

 

specifically at 3:7 or 2:8 to leave no residue on the lens). Be extremely careful when wiping the  

 

lens clean. Do not, under any circumstances, use any sort of corrosive solution to clean the lens. 

 

Under no circumstances should dry a cotton swab, dry cloth or dry lens cloth be used to wipe the  

 

lens as this will cause scratches and damage the surface of the lens. Do not use water to clean  

 

the lens, as this will leave a residue on the lens surface, as well as possibly resulting in the  

 

formation of mildew. 

 

The light barrier and glass ring in the carriage and the frosted glass must be kept clean. If oil  

 

stains or dust appear on the surfaces, use a cotton swab dipped in mixture of pure alcohol and  

 

ether to carefully clean as such contamination will affect observation results using transmitted  

 illumination. 

 

10.4 Moving the Microscope 

 

Try to move the microscope as little as possible. 

 

If it is absolutely necessary to move the microscope, be sure the eyepieces, optical body, focusing  

 

mechanism and the adjustment rings are secured in place before moving the microscope. As well,  

 

be sure the power is turned off and the cord unplugged.  

 

When moving the microscope use both hands, one under the base and the other firmly holding the  

 

top of the vertical bracket. 

 

The microscope should be kept in an upright position the entire time it is being moved. 

Содержание 1100200600841

Страница 1: ......

Страница 2: ...the mechanical as well as optical components of our microscopes Therefore all descriptions and illustrations in this instruction manual including all specifications are subject to change without prio...

Страница 3: ...171B 8 4 3 Binocular Observation Tubes GM 161B 9 4 4 Trinocular Observation Tubes GM 171T GM 161T 9 4 5 Microscope Carrier 9 4 6 Base 10 4 7 Illumination 13 4 8 Stage 14 5 Accessories 15 5 1 Gemstone...

Страница 4: ...bulb 22 7 5 Microscope operation 25 8 Camera and video equipment installation 26 8 1 Camera installation and use 26 8 2 Video installation and use 26 9 Operation safety measures 27 10 Microscope main...

Страница 5: ...e with the local power supply Using the microscope with a non specified voltage rating can lead to damage of the instrument 1 5 When changing the bulb please note Only specified halogen bulbs and fluo...

Страница 6: ...5 2 NOMENCLATURE 2 1 GM 161B Fig 1...

Страница 7: ...6 2 2 GM 171T Fig 2...

Страница 8: ...fferent eyepieces and objective lenses changing the total zoom range from 2 25X to 200X of GM 171 and 2 25X to 180X of GM 161 make this microscope a welcome addition to any laboratory The instrument s...

Страница 9: ...both sides of the optical body rotate these knobs to change the magnification strength For the specific magnification strength see the values on the adjustment knob the current magnification strength...

Страница 10: ...50 of the lights will enter eyepieces and 50 of the lights for camera The brightness could be fully satisfied with the board band anti reflection technology and high brightness in illumination At the...

Страница 11: ...will keep spinning in order to avoid excessive force being exerted on the gears 4 6 Base The tilting axis on the base allows users to angle the microscope arm bracket microscope carrier and optical bo...

Страница 12: ...tch is located on the right side of the base Rotate the dial to turn the transmitted illumination on and off and adjust intensity Fig 12 On the underside of the base is located a circular rubber turre...

Страница 13: ...12 Fig 12 Fig 13...

Страница 14: ...ight diffuser dish carriage light barrier and 41mm 2mm adjustable diaphragm Turn the main power on Fig 15 rotate the adjustment dial on the right side of the base to turn illumination on Fig 16 rotate...

Страница 15: ...the incident illumination power switch located on the lamphouse to turn on and off Fig 20 Adjust the gooseneck mount to achieve desired incident angle Fig 19 Fig 19 Fig 20 4 8 Stage The stage contains...

Страница 16: ...of the holder as this may cause the wires to bend permanently damaging the instrument Fig 21 5 2 Polariser The polarising unit consists of a polariser and an analyser with holding screw and is general...

Страница 17: ...eaded bolt is facing the incident illumination source while the handle of the spring loaded bolt is facing the microscope arm Fig 25 Use jewel tweezers to insert a diamond into the clamp so the pointe...

Страница 18: ...te objective according to working distance and total magnification requirements Screw on the desired objective After the objective has been installed the working distance will be changed and microscop...

Страница 19: ...al camera supplies shop The same method can be applied to connect a digital camera to the microscope 5 9 C mount camera adapter The c mount camera adapter is inserted into the top of the trinocular tu...

Страница 20: ...plate can be used directly on top of the diaphragm on the stage for samples too large for the standard gem holders The plate comes equipped with a black white plate and a frosted glass plate Choose t...

Страница 21: ...When moving the various components especially the optical components fingers and oily materials should not come in contact with the surface of the lenses Fingerprints and oil stains on the lenses wil...

Страница 22: ...ar the interpupillary distance is too large If two overlapping fields of view appear the interpupillary distance is too small Fig 30 Fig 31 7 2 Focus and diopter adjustment Adjust the height of the fo...

Страница 23: ...stment knob until the desired magnification is achieved If necessary the eyepieces may be changed or an additional objective lens may be installed see section on additional objectives installation at...

Страница 24: ...at the bottom of the heat dispersal Fig 34 unit and open the panel to change the bulb Fig 35 and Fig 36 Once the bulb has been changed close the panel retighten the screw Fig 37 and return the microsc...

Страница 25: ...of the housing and while pulling in from the backside Fig 40 pull outwards to remove the socket Fig 41 Remove the bulb from the socket Fig 42 replace with a new one push the socket back into the lamp...

Страница 26: ...mode If further accessories are necessary for observation install the required accessories and begin observation again according to the aforementioned steps At this point the angle of the microscope...

Страница 27: ...oosened in the second step 8 2 Video installation and use The video device includes a video camera a c mount camera adapter and a camera port Among which the camera port is supplied by Motic while a v...

Страница 28: ...ould be kept away from dust water water vapor and corrosive materials If dust or water works their way into the microscope it may result in mildew growth damaging the instrument Please be aware that o...

Страница 29: ...circumstances use any sort of corrosive solution to clean the lens Under no circumstances should dry a cotton swab dry cloth or dry lens cloth be used to wipe the lens as this will cause scratches and...

Страница 30: ...roscope Maintenance and Care section of this manual Low definition or unclear image 1 Unable to bring specimen into focus see section above for possible reasons and solutions 2 Objective lens may be d...

Страница 31: ...30 APPENDIX A GM 171 OPTICAL DATA Note WD Working Distance Mag Magnification FD Field Diameter Note High eyepoint eyepiece...

Страница 32: ...31 APPENDIX B GM 161 OPTICAL DATA Note WD Working Distance Mag Magnification FD Field Diameter Note High eyepoint eyepiece...

Страница 33: ......

Отзывы: