10
Fig.5
Fig.6
Fig.7
Fig.8
If, when trying to focus, the body and frame keep sliding downwards adjust the coarse focus
torque adjustment rings to tighten the mechanism.
The focus travel range is 92mm. The focusing knobs are equipped with a slip control function,
which means when the end of the travel range is reached the knobs will keep spinning in order to
avoid excessive force being exerted on the gears.
4.6 Base
:
The tilting axis on the base allows users to angle the microscope arm (bracket, microscope carrier
and optical body) from 0° ~ 45° as observation requires (Fig.9). Adjust the screws located inside
the axis where the arm meets the base to loosen or tighten the rotating action (Fig.10).
Installed within the base are the illumination switching circuits.
The 2.5A (labeled T2.5A) fuse and MAIN power switch are located on the backside of the base.
(Fig.11)