21
4.10 Polarizing Observation
Insert the polariser (marked with
“P”
) into the slot.
Insert the Analyser (marked with
“A”
) into the slot.
Rotate the analyser according to demands.
(Fig.20)
Note:
1.
Before inserting Polariser and Analyser in slot please make sure the filter surface is dust-free,
for cleaning method please refer to
item 7. “CARE AND MAINTENANCE”
2.
Before starting to observation, please make sure
item 4.6
has been set up properly
4.11 Brightness and Contrast Adjustment
Blue filter is used for increasing color temperature in routine bright field microscopy
Frosted filter reduces irregularity in the illumination field, but also reduces the brightness.
To ensure enough brightness and better image quality, remove the frosted filter out of light path
when using the high magnification objective and low reflectivity of sample.
For the best contrast and image quality, adjust the condenser aperture diaphragm accordingly
when the objective changed.
4.12 Filter selection and installation way
Filter type
Procedure
ND (Neutral Density) filter
For intensity adjustment in photomicrography
Blue filter
For general microscopy and colour
photomicrography