M2118
Operating Manual
V1.06 © 17.10.2011 MOSTEC AG, CH-4410 Liestal
4
2.Adjust the limit contacts:
1. Push and hold the button <ENTER> for 4 seconds.
ð
The display will show up the program menu.
2. Change to menu SP_1 for limit contact #1 or SP_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
ð
The display shows the actual value of the limit contact.
3. Change the value for the limit contact by the button <+> or <->, then press <ENTER>.
4. Change to menu HSt_1 for limit contact #1 or HSt_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
5. Change the value for the hysteresis by the button <+> or <-> , then press <ENTER>.
6. Change to menu LEd_1 for limit contact #1 or LEd_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
7. Change the lamp status
by the button <+> or <-> , then press <ENTER>.
"nor": LED lamp ON, when input > setpoint
"Inr": LED lamp ON, when input < setpoint
8. Change to menu rEL_1 for limit contact #1 or rEL_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
9. Change relay operating mode
by the button <+> or <-> , then press <ENTER>.
"nor": Relay ON, when input > setpoint (normal mode)
"Inr": Relay ON, when input < setpoint (fail-safe mode)
10. Press button <ESC>
ð
the display shows the actual value.
C. Adjusting the limit contacts with programming software
C
onnect the link cable with Adapter to the Computer and the display. (see I, page 6)
1. start "M
pro
"
2. Choose "M2118" in menu "Instrument" <A>
3. Choose "Linear conversion table" under "Range" <B>
4. Choose "Adjust the limit contacts" <C>
5. Enter the values for the limit contacts and all other settings (hysteresis etc.)
6. Choose "Program the limit contacts"
D. Adjusting the display range with programming software
C
onnect the link cable with Adapter to the Computer and the display. (see I, page 6)
1. start "M
pro
"
2. Choose "M2118" under "Instrument" <A>
3. Choose "Linear conversion table" under "Range" <B>
4. Enter the value for input range, output range and decimal place
5. Choose "Program"