Oxygen Sensor Exchange / Repair
SENSOR EXCHANGE:
Sensor is packed with its cover to keep its electrolyte in passive mode. Open the sensor cover just
before 24 hours of installation and keep it at ambient condition that will operate. Passive electrolyte
will get active with this way and start to measure correctly after installation into the OPT100 panel. Do
not forget to note sensor installation date onto the panel. Please check SENSOR CALIBRATION part
for new sensor calibration.
Sensor date code:
②⑦①③
Week
Year (2013)
SENSOR REPAIR:
Do not open the sensor for any repair attempt. There isnt any repairable part inside the
sensor and it will be damaged permanently. Inner Electrolyte can also create severe irritation
on skin.