7.
Span ve Zero Parameter Adjustment
Press button for 5 seconds.
PRG
11
o2
Code
122
OPT100 will request safety code. Write 122 with arrow buttons and press SET.
SET
CaLb
6rvb
Calibration adjustment parameter will be opened.
SET
Zero
11
ZERO parameter screen will be opened. Decrease the value to 0,0 with arrow
button via pressing it continuously. Be sure that Calibration fitting is connected to
the sensor and N2 bottle valve is opened with 99,9% N2 gas supply to the sensor.
After reaching to 0 value press SET button and stop the gas supply.
SET
span
209
Normal Operation Screen
SPAN parameter shows ambient O2 value. Apply calibration for values different
than 20,9 (lower or higher) Reach 20,9 value by pressing arrow buttons.
After reaching 20,9 value do not touch any button. OPT100 will return to normal
operation mode by itself in 30 seconds.
209
o2
Normal Operation Screen.
7