SECTION 3
27
In addition to manipulating
SPEED, GAIN,
and
FILTER
, the user can toggle different lead groups during printing
by selecting
LEADS
. The change in lead groups is apparent on the printout whereas the waveform display will
remain in the default display of 2.5 seconds of Leads I, II, and V1-V6.
During 3-channel rhythm printing, the available lead groups are:
1.
Default (user-selected in configuration)
2.
I-II-III
3.
aVR, aVL, aVF
4.
V1-V2-V3
5.
V4-V5-V6
During 6-channel rhythm printing, the available lead groups are:
1.
Default (user-selected in configuration)
2.
I-II-III-aVR-aVL-aVF
3.
V1-V2-V3-V4-V5-V6
During rhythm printing the user can place the writer in Standby mode by pressing
STOP
. To continue rhythm
printing for the same patient without advancing to a new page, select
RHY
.
Transfer to a USB Memory Stick
The user can transfer all the ECG records from the ELI 230 to an external USB memory stick at any time. Using a
PC, create a directory on the USB memory stick called “Records.” When done, plug the USB memory stick into the
USB port on the ELI 230 (same port as used for the AM12 acquisition module). Once the USB memory stick is
connected, the device will automatically start transferring all existing records to the USB memory stick. The device
will display the message “ECG TRANSFER IN PROGRESS”. Upon completion another message will display
confirming the number of records transferred, “TRANSFERRED: 20/20, DONE – remove USB stick”. The device
will automatically return to the real-time ECG display when the USB memory stick has been removed. All
successfully transferred ECG records will have been cleared from the device.
NOTE: Not all USB memory sticks may fit the device’s USB port.
NOTE: If the device appears to be unresponsive, hold the Off key for 10 to 15 seconds to reboot.
Deleting Records
To delete all existing patient records from the ELI 230:
•
Select
MORE
, followed by
CONFIG
.
•
Select application
4. Service
followed by
Yes
.
•
Select application
3. Clear all ECG Records
. Once selected, a series of dots will start to appear on the
display. This signals to the user that the records are being deleted. Once the deletion process is completed,
the display will return to the main configuration screen.
•
Select
DONE
to return to the real-time ECG display.