SECTION 3
25
Best 10 Seconds Selection
The ELI 230 incorporates a one-minute memory buffer for collection of ECG data. When the Best 10 feature is
enabled, the device will automatically select the best 10 seconds of ECG from within the one-minute buffer. The
best 10 seconds is determined based upon measures of high and low-frequency noise found in the 10-second ECG
segments. Should a single limb lead or two precordial lead fail conditions occur, the Best 10 feature is disabled until
the limb lead fail or precordial lead fail condition is resolved. At resolution, the Best 10 feature becomes available
and the selection buffer continues.
Users can switch between BEST 10 or LAST 10 by selecting
BEST10 or LAST10
. Switching the view will
automatically generate a second printout if auto print is selected. Otherwise the clinician will be required to select
.
Pacemaker Spike Notification
Turn the pacemaker spike notification marker on or off by selecting Yes or No in the configuration menu. Selecting
Yes will cause a spike location marker to display at the bottom of the printout.
Printing
If Auto-Print is enabled in the configuration, an ECG is printed following acquisition. To print a manual printout,
select
.
If the auto-print configuration is disabled, a 10-second preview will assist in ensuring a quality ECG acquisition
prior to printing. When you acquire an ECG, the electrocardiograph captures a minimum of the last 10 seconds.
The relationship between the display and the printout is the same – what is displayed in the ECG acquisition view is
what will be printed.
In order to change the speed, gain, filter, or print format in the acquired ECG view, select
MORE
. To manipulate
the print format of the acquired ECG regardless of the plot format configuration setting, select
FMT
. Continue to
press
FMT
until the desired print format is displayed, then select
BACK
. The display will return to acquired ECG
view. Select
to make an ECG printout copy in the new plot format; select
DONE
to return to the real-time
ECG view.
Storage
The ELI 230 will automatically store up to 20 ECGs in its internal memory; however, the ECG records must be
transferred via USB memory stick to a PC running ELI Link to be reviewed. Starting with 20 and counting down to
0, the device will display the number of remaining storage slots available just below the display sweep speed on the
right side of the display in acquired ECG view. Once the device has stored 18 records (2 storage slots remaining), a
red box will appear around the countdown number. If set to Yes in the configuration menu, a storage warning
message will appear in the center of the display on real-time ECG view to notify the user that maximum storage is
about to be reached. Once maximum storage is reached, the device will automatically delete the oldest record to
make room for the new record (first in, first out). The user can also enter the Service menu and delete all patient
records at any time, or transfer all the ECG records to a USB memory stick. (See
Transfer to a USB Memory Stick
.)
Содержание ELI 230
Страница 14: ...GENERAL CARE xiv...
Страница 23: ...SECTION 1 3 ELI 230 System Illustration Figure 1 1...
Страница 24: ...SECTION 1 4 ELI 230 Left Side Figure 1 2 ELI 230 Rear Figure 1 3...
Страница 25: ...SECTION 1 5 ELI 230 Base Figure 1 4...
Страница 40: ...SECTION 2 20...
Страница 48: ...SECTION 3 28...