SECTION 6
36
Once SIM card is in place, replace the cover and the Phillips screw. Power on the ELI 10 Rx and adjust the
appropriate menus for GSM/GPRS mobile operation.
Worklist Management
The ELI 10 Rx can download and process ECG order
lists from the E-Scribe or another compatible information
management system which identifies the ECGs (or ECG orders) needed for particular patients. Implementation of
an order based workflow can significantly reduce demographic data entry errors at the electrocardiograph. Orders
are deleted from the list when the ordered ECG is acquired.
When Worklist Management is set to Standard, new order lists are appended to the remaining list. When set to
Refresh, each new order list will override the previously downloaded one.
Comm. Protocol
This device may be set to a communication protocol of either DICOM or UNIPRO. Set to UNIPRO to transmit to
E-Scribe or ELI LINK; set to DICOM to transmit to a DICOM compatible management system.
NOTE: This parameter must be entered under the direction of the IT Manager of the facility where the
device is installed.
Owner Name
Option currently not available.
Multi-Protocol Configuration
Site Number (Second and Third site as well)
The ELI 10 Rx allows for the configuration of up to three site numbers in order to manage multiple studies. A site is
defined as a location from where an ECG is transmitted. In a research environment, it is recommended that site
numbers designate the investigator’s site for ECG records stored in an E-Scribe Rx data management system, and
must be defined for transmitting ECGs to the data management system. Up to four digits can be used for the site
number; numbers from 0 – 4095 are supported
NOTE: If the second or third site number fields are blank, a site is not identified. The number “0” is a
valid site number.
NOTE: Adding, editing, or deleting any site number from the configuration will cause the unit to default to
short ID format for that site number until a new custom ID format (see Section 4) is downloaded for the
new site number.
Site Name (Second and Third Site As Well)
Defines your investigator site name. You can enter up to 30 alphanumeric characters. The site name prints in the
middle of the bottom edge of the ECG printout at the E-Scribe.