Installation(Continued)
1.
Drain Connection
The Maestro is equipped with 1.5" Diameter drain and connections. DO NOT modify the
drain
line or connections without contacting MOPEC first. The units are built in accordance
with
the rough-in drawings specific to the unit. The drain connection provided unless
specifically
requested otherwise is Orion Acid Waste Pipe 1.5" Diameter. The height is per
the rough-in
drawings provided at the time of the order.
2.
Water supply Connection
The water supply connection are 1/2" Diameter copper pipe. The unit has been cleared of
water and debris during the manufacturing and testing of the unit. The rough-in drawing for
the unit suggests a 1/4" turn valve to be connected to the wall. This is to allow the water to
be localized and disconnected should the need arise to work on plumbing in the future.
Before connecting the unit to the facility water supply, check the lines in the facility for debris
by flushing the lines for a moment. After the connections are made remove the aerator from
the faucet and turn the water on and open the faucet and other water options after the faucet
to allow any debris to evacuate the system. Then one option at a time, open the valves for
the other water options(i.e. Perimeter Rinse, Spray Hose, and then the Disposal). This is the
time to look for leaks or other plumbing issues. The plumbing is tested for leaks and
functions at the facility and is put under pressure for 45 Minutes to check for leaks. With the
maestro pex plumbing, if any leak issues arise, find the leak and replace the line.
3.
Ventilation Connection
Ventilation is one of two types for the Maestro; In House Ventilation or Recirculation. For in
house the unit is connected to the facility ventilation system via duct work. The duct work will
vary depending on the model of the Maestro.
a.
In House
If your unit elevates, it will have two 8" diameter flexible duct hoses. The duct hoses
are connected to the grossing station chamber door on top of the unit and to
corresponding stubs at the ceiling per the rough-in drawing. The ducts are provided
as well as the stainless steel band clamps. The typical flexible duct hose length is 3
foot.
b.
Recirculation
For recirculation models, the filters will need to be installed and the clearance above
the unit verified at its highest elevation. You should have a minimum of 12 inches or
more above the Maestro at its highest position. Air flow is generated by multiple fans
in the recirculation models. They are set up so one fan is on constantly when the unit
is on and the other fan is controlled by a rheostat to control volume/speed.