background image

Voltage, Terminal V: 

DC Range 

AC Range  Resolution 

Noise Floor 

Accuracy 

(%+counts) 

600 V 

600 V 

275 µV 

9 mV 

0.5 + 20 

600 V 

430 V 

140 µV 

5 mV 

0.5 + 20 

455 V 

320 V 

95 µV 

3 mV 

0.5 + 20 

365 V 

255 V 

70 µV 

3 mV 

0.5 + 20 

260 V 

185 V 

50 µV 

2 mV 

0.5 + 20 

200 V 

140 V 

35 µV 

1090 µV 

0.5 + 20 

135 V 

95 V 

25 µV 

730 µV 

0.5 + 20 

60 V 

45 V 

20 µV 

610 µV 

0.5 + 20 

40 V 

30 V 

10 µV 

305 µV 

0.5 + 20 

30 V 

22 V 

6.4 µV 

205 µV 

0.5 + 20 

25 V 

18 V 

4.8 µV 

155 µV 

0.5 + 20 

18 V 

12 V 

3.2 µV 

105 µV 

0.5 + 20 

14 V 

10 V 

2.4 µV 

80 µV 

0.5 + 20 

9 V 

6 V 

1.6 µV 

55 µV 

0.5 + 20 

1000 mV 

700 mV 

300 nV 

10 µV 

0.5 + 20 

670 mV 

475 mV 

150 nV 

5 µV 

0.5 + 20 

500 mV 

350 mV 

100 nV 

4 µV 

0.5 + 20 

400 mV 

275 mV 

75 nV 

3 µV 

0.5 + 20 

280 mV 

200 mV 

50 nV 

2 µV 

0.5 + 20 

220 mV 

150 mV 

40 nV 

2 µV 

0.5 + 20 

150 mV 

100 mV 

25 nV 

1 µV 

0.5 + 20 

 

Current, Terminal A: 

DC Range  AC Range  Resolution  Noise Floor 

Accuracy 

(%+counts) 

10 A 

9 A 

3.8 µA 

120 µA 

0.5 + 50 

8.5 A 

6 A 

1.9 µA 

60 µA 

0.5 + 50 

6 A 

4.5 A 

1.3 µA 

40 µA 

0.5 + 50 

5 A 

3.5 A 

1 µA 

30 µA 

0.5 + 50 

3.5 A 

2.5 A 

0.7 µA 

20 µA 

0.5 + 50 

2.75 A 

2 A 

0.5 µA 

15 µA 

0.5 + 50 

1.75 A 

1.25 A 

0.4 µA 

10 µA 

0.5 + 50 

 

Resistance 

Range 

Resolution 

Noise Floor 

Accuracy 

(%+counts) 

9 MΩ

 

1.5 Ω

 

50 Ω

 

1.0 + 50 

8 MΩ

 

1 Ω

 

35 Ω

 

1.0 + 50 

6 MΩ

 

750 mΩ

 

25 Ω

 

1.0 + 50 

4 MΩ

 

500 mΩ

 

20 Ω

 

1.0 + 50 

3 MΩ

 

375 mΩ

 

12 Ω

 

1.0 + 50 

2 MΩ

 

250 mΩ

 

8 Ω

 

1.0 + 50 

9.6 kΩ

 

1.5 mΩ

 

48 mΩ

 

1.0 + 50 

8.2 kΩ

 

1 mΩ

 

32 mΩ

 

1.0 + 50 

6.1 kΩ

 

750 µΩ

 

24 mΩ

 

1.0 + 50 

4.1 kΩ

 

500 µΩ

 

16 mΩ

 

1.0 + 50 

3 kΩ

 

375 µΩ

 

12 mΩ

 

1.0 + 50 

2 kΩ

 

250 µΩ

 

8 mΩ

 

1.0 + 50 

 

 

Содержание Mooshimeter BLE-DMM-2X-01A

Страница 1: ...MOOSHIMETER USER S MANUAL Mooshim Engineering LLC 2565 3rd St Suite 315 San Francisco CA 94107 August 2014 Rev 1 2014 Mooshim Engineering All Rights Reserved...

Страница 2: ...defines a level of safety appropriate for the transient overvoltages present in fixed equipment installations at the distribution level This includes the mains installation of a building distribution...

Страница 3: ...least CAT III 600V Keep fingers behind the guards on the leads Use caution working with voltages above 30VAC rms 42VAC peak or 60VDC Voltages this high pose risk of shock Never apply more than 600V b...

Страница 4: ...C Common terminal for all measurements A Input terminal for measuring currents to 10A RMS AC and DC V Input terminal for measuring voltages to 600V RMS AC and DC Input terminal for measuring resistan...

Страница 5: ...he circuit before assuming the circuit is safe Two of the Mooshimeter s input terminals are capable of making basic voltage measurements The V terminal can measure up to 600V RMS and is intended for t...

Страница 6: ...n determining whether a circuit is live if the meter does not report a current This may indicate either a poor connection or a blown fuse NOTE AC current range is limited by the instantaneous current...

Страница 7: ...surements The terminal can measure small voltages with a resolution of as fine as 25nV per count and a noise floor as low as 1 V This can be used to with external current shunts to measure currents ov...

Страница 8: ...ay engage the internal protection circuitry Should this occur wait 5 minutes before taking further measurements Resistance is measured between the and C terminals Polarization is not important The met...

Страница 9: ...nal protection circuitry Should this occur wait 5 minutes before taking further measurements To test diodes or other silicon junction devices a test current is pushed through the device under test and...

Страница 10: ...standard BNC adapter Measurements using the A or terminals can be used with the C terminal as usual To use a BNC adapter with the V input use the A terminal as the common input The A and C terminals a...

Страница 11: ...en finds the slope to calculate resistance This can be done with either the internal current shunt or an external current clamp Using a current clamp allows the measurement to be done without breaking...

Страница 12: ...urement requires voltage and current measurements Current measurement can be done with either internal or external current shunt mode Power Factor is a unitless measure so the gain accuracy of the mea...

Страница 13: ...ning the meter 1 Disconnect all test leads 2 Unscrew both retaining screws from the bottom of the case with a Phillips head screwdriver 3 Separate the two halves 4 Taking special care not to bend the...

Страница 14: ...5 Your Mooshimeter is now fully disassembled The batteries fuse and reset button are now accessible...

Страница 15: ...shock Do not operate the meter while partially disassembled 1 Place the PCB in the bottom enclosure half Use the J shaped fin for alignment with the corresponding cut out in the PCB 2 Close the two ha...

Страница 16: ...ty the Mooshimeter s LED will blink slowly several times before turning off If the light does not blink check the battery polarity 5 Close the Mooshimeter per directions above CLEANING Gently wipe the...

Страница 17: ...ter is not accurate below 1 The actual resistance of the fuse should be 20m FIGURE 8 FUSE TEST FUSE REPLACEMENT WARNING Only use appropriately rated replacement fuse 1 Open the Mooshimeter per directi...

Страница 18: ...er SD cards or cards formatted differently cannot be written to 1 Open the Mooshimeter per the directions above 2 Press the SD card further into the slot until you feel a click 3 Release The SD card w...

Страница 19: ...IMETER WARNING To avoid electric shock disconnect all leads from the meter 1 Open the Mooshimeter per the directions above 2 Press and Hold the button shown for 5 seconds 3 Close the Mooshimeter per t...

Страница 20: ...ing Altitude Up to 2000m Maximum Temperature Range 5C to 40C Nominal Temperature Range 18C to 28C Derate accuracy outside of this range Input Power 3V 100mA from installed batteries Battery 2x AA Alka...

Страница 21: ...0 5 20 500 mV 350 mV 100 nV 4 V 0 5 20 400 mV 275 mV 75 nV 3 V 0 5 20 280 mV 200 mV 50 nV 2 V 0 5 20 220 mV 150 mV 40 nV 2 V 0 5 20 150 mV 100 mV 25 nV 1 V 0 5 20 Current Terminal A DC Range AC Range...

Страница 22: ...0 nV 2 V 0 5 20 220 mV 150 mV 40 nV 2 V 0 5 20 150 mV 100 mV 25 nV 1 V 0 5 20 Diode Voltage Drop Range Resolution Noise Floor Accuracy 2 2 V 300 nV 10 V 1 0 50 1 2 V 150 nV 5 V 1 0 50 825 mV 100 nV 3...

Страница 23: ...ons rules This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must a...