IPF
Page 3
Model/Serial Number.
Moore Industries uses a
system of individual unit model and serial numbers to
keep track of factory configuration and options for
IPF’s shipped or serviced.
If service information or assistance is required for
your IPF(s), be sure to provide the factory with the
unit model and serial numbers to assist our profes-
sional technicians in their effort to give you prompt,
efficient service.
The IPF serial and model numbers can be found on
a metal tag on the side of the unit housing. The
example outlines the significance of each field of
information in a typical IPF model number.
Calibration
Every IPF is fully tested and calibrated at the factory
prior to shipment. Before installation, however, your
IPF’s should be bench checked to verify proper
operating levels, and to set the desired unit zero and
span.
Calibration should be conducted in an appropriate
testing environment. By carrying out the procedures
at a technician’s bench or similar lab-type setup, any
unit damage that may have occurred during ship-
ment can be discovered safely, i.e., isolated from the
intended application.
Unit Type
Input
Output
Supply Pressure
Options(s)
Housing
IPF / 4-20MA / 3-15PSIG / 20PSI / -FR1 [ WP ]
EXAMPLE