Computer Reference
7.8 Menu 99: Service menu (hidden)
1 Adjust pot. (potentiometer)
If any service regarding the pendulum has been done the correct potentiometer
position can be checked here.
Follow display instructions.
2 Calibrate pot.
After adjusting pot. a calibration shall be performed here. Also after setting to
default in the service menu a calibration has to be done here.
Follow display instructions.
3 Set to default
Erases the memory. Formats and sets some data to default.
NOTE! A calibration must be performed after this action.
4 Printer
Form feed:
Line feed:
Page setting:
Lines and margins can be set here.
Print report:
To check printer connection.
5 Motor
Motor current limit.
Motor forward, backward and stop.
6 Service setup
Pedal ref.:
Metronome setting
Pedal min.:
Lowest pedal rpm to obtain workload on the bike can be
set here.
Default setting is 30 rpm.
Cykelkonst.:
This value can be adjusted so that the power on the bike
can be corrected to the pedals or to the flywheel.
NOTE! In the Astrand test the power is calculated on the
flywheel. Default setting is 1.00 which means there is no
extra adjustment done for transmission loss. If for example
setting is 1.04 this means that loss is calculated to 4%.
Max. accepted
pulse deviation:
This figure is the deviation allowed during steady state to
calculate a test.
Default setting <5> bpm.
Weight unit:
0 = kg, 1 = pounds
Distance unit:
0 = km, 1 = miles
Opt. pulse:
on/off
Opt. metronome: on/off
Содержание 839 E
Страница 1: ...Manual English Monark 839 E Ergomedic 839 E...
Страница 2: ......
Страница 17: ...Monark Ergomedic 839 E 17 Computer Reference 7 1 6 Menu system 3 4...
Страница 18: ...Monark Ergomedic 839 E 18 Computer Reference Kommandotyp 0 Terminal PC 1 3 Se manual 0...
Страница 30: ...Monark Ergomedic 839 E 30 Command type 0 Terminal PC 1 3 Se manual 0 Computer Reference...
Страница 40: ...S 780 50 Vansbro Sweden Tel 46 0 281 59 49 40 Fax 46 0 281 719 81 www monarkexercise se...