Docon
®
7
Product description
Page 21 of 74
Disassembly of the Docon 7 sealing handle
To clean the electrodes, firstly remove the blue lever and the electrode (see
Figure 6
Figure 7
Figure 8
Figure 9
1. Press the lever as described in
against the handpiece until the two elec-
trodes in the handle head come into contact.
2. Press the plunger pins in on both sides until an audible "click" is heard (see
Figure 7 and
). Use a ballpoint pen or similar to push in the plunger pins.
3. Push the electrode and lever upward and remove (see
The round handpiece and electrodes should be cleaned with a lint-free cloth.
The removed lever with electrode can be washed under running water.
Dry all parts carefully. Ensure that the electrodes are completely dry to avoid
spark discharge.
Plunger pins
Plunger pins