15
Molift QuickRaiser 205 / www.etac.com
Maintenance
Cleaning and disinfection
Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the lifter. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products.
• Remove lifting arm cover to clean underneath
• If nessesary, remove the control panel cover to
clean inside
• If nessesary, remove the leg support and clean
it with running water
• Remove hair and pile from the castors and
verify that the castors rotates freely.
Clean contact and hand control with isopropyl
alcohol to remove grease and dirt if necessary.
Check emergency stop after cleaning.
The lift should not be exposed to running water.
Make sure not to damage or
remove labels when cleaning.
Reconditioning
Follow cleaning and assembly procedure, complete
periodic inspection and use checklist after assem-
bly to recondition the lifter.
Recycling
Refer to “Recycling instructions” for how to prop-
erly dispose of product. See www.etac.com.
Spare Parts
A list of spare parts is available on request.
Periodic Inspection Scope
Periodic inspection is a visual examination (particu-
larly of the lifter’s load bearing structure and lifting
mechanism with attachments, brakes, controls,
safety devices and person-support devices)
according to Periodic Inspection Report for Molift
QuickRaiser 205.
See www.etac.com.
In the event of danger to safety,
the lifter shall immediately be
taken out of service and marked
clearly with “out of order” and
shall not be used until the lifter is
repaired
Periodic Inspection shall be performed at least
once a year or more frequently if required by local
requirements. The inspection must be performed
by service personnel authorized by Etac. Contact
Etac at [email protected] for training and
authorization or recommendation of an approved
service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report for
Molift QuickRaiser 205. The reports should be
retained by the person(s) responsible for servicing
the lifter. If the inspection reveals defects and
damages, the lifter shall be taken out of operation
and marked “out of order” until it is repaired.
After performing periodic inspection reset the
service indicator
Reset service indicator
Reset the service indicator by pushing in and
holding “up” and “down” on the control panel for
5 seconds. A green light will blink for 5 seconds,
imedietly after this light starts to blink push in and
hold both “in” and “out” for five seconds. You
will hear two short beeps. The service light will no
longer flash. It will start to flash again when one
year has passed and it is time for new periodic
inspection.
Periodic Inspection